用于在射频偏置电极处测量离子能量分布的减速场分析仪。

Retarding field analyzer for ion energy distribution measurements at a radio-frequency biased electrode.

作者信息

Gahan D, Dolinaj B, Hopkins M B

机构信息

National Centre for Plasma Science and Technology, Dublin City University, Glasnevin, Dublin 9, Ireland.

出版信息

Rev Sci Instrum. 2008 Mar;79(3):033502. doi: 10.1063/1.2890100.

Abstract

A retarding field energy analyzer designed to measure ion energy distributions impacting a radio-frequency biased electrode in a plasma discharge is examined. The analyzer is compact so that the need for differential pumping is avoided. The analyzer is designed to sit on the electrode surface, in place of the substrate, and the signal cables are fed out through the reactor side port. This prevents the need for modifications to the rf electrode--as is normally the case for analyzers built into such electrodes. The capabilities of the analyzer are demonstrated through experiments with various electrode bias conditions in an inductively coupled plasma reactor. The electrode is initially grounded and the measured distributions are validated with the Langmuir probe measurements of the plasma potential. Ion energy distributions are then given for various rf bias voltage levels, discharge pressures, rf bias frequencies--500 kHz to 30 MHz, and rf bias waveforms--sinusoidal, square, and dual frequency.

摘要

研究了一种用于测量等离子体放电中撞击射频偏置电极的离子能量分布的减速场能量分析仪。该分析仪结构紧凑,因此无需差动抽气。该分析仪设计为放置在电极表面,取代衬底,信号电缆通过反应器侧端口引出。这避免了对射频电极进行修改——而内置在这种电极中的分析仪通常需要进行修改。通过在电感耦合等离子体反应器中在各种电极偏置条件下进行实验,展示了该分析仪的性能。电极最初接地,测量的分布通过等离子体电位的朗缪尔探针测量进行验证。然后给出了各种射频偏置电压水平、放电压力、射频偏置频率(500 kHz至30 MHz)以及射频偏置波形(正弦波、方波和双频)下的离子能量分布。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验