Lüthy R
Infection. 1977;5(2):115-6.
The urease method developed by Noone et al. for rapid bioassay of aminoglycoside antibiotics in serum is described in detail. The accuracy of the method was improved by using a BM 1-09 electrode assembly (Metrohm, CH-9100 Herisau, Switzerland) in conjunction with a digital pH-Meter (Metrohm E 500). The mean difference between spiked serum samples and measured concentrations was +0.37 +/-0.78 microgram/ml. The coefficient of determination between the urease and agar diffusion method was 0.94. Disadvantages of the method are a low sensitivity at concentrations below 1.5 microgram/ml and a large serum sample (1.5 ml). Advantages are its simplicity and rapidity.