Lüthy R
Infection. 1977;5(2):115-6.
The urease method developed by Noone et al. for rapid bioassay of aminoglycoside antibiotics in serum is described in detail. The accuracy of the method was improved by using a BM 1-09 electrode assembly (Metrohm, CH-9100 Herisau, Switzerland) in conjunction with a digital pH-Meter (Metrohm E 500). The mean difference between spiked serum samples and measured concentrations was +0.37 +/-0.78 microgram/ml. The coefficient of determination between the urease and agar diffusion method was 0.94. Disadvantages of the method are a low sensitivity at concentrations below 1.5 microgram/ml and a large serum sample (1.5 ml). Advantages are its simplicity and rapidity.
详细描述了Noone等人开发的用于血清中氨基糖苷类抗生素快速生物测定的脲酶法。通过将BM 1-09电极组件(万通公司,瑞士黑里绍CH-9100)与数字pH计(万通E 500)结合使用,提高了该方法的准确性。加标血清样品与测定浓度之间的平均差异为+0.37 +/-0.78微克/毫升。脲酶法与琼脂扩散法之间的决定系数为0.94。该方法的缺点是在浓度低于1.5微克/毫升时灵敏度较低,且血清样品量较大(1.5毫升)。优点是操作简单、快速。