Zahedi E, Jaafar R, Ali M A Mohd, Mohamed A L, Maskon O
Biomedical Engineering Laboratory, School of Electrical Engineering, Sharif University of Technology, Tehran, Iran.
Physiol Meas. 2008 May;29(5):625-37. doi: 10.1088/0967-3334/29/5/008. Epub 2008 May 7.
This study was conducted to investigate the utility and efficacy of finger photoplethysmogram pulse amplitude (PPG-AC) in comparison with the standard Doppler ultrasound in assessing an endothelial function via flow-mediated dilation (FMD). High-resolution B-mode scanning of the right brachial artery (BA) of 31 healthy subjects aged 39.7 +/- 11.3 (range 22-64) years and 52 risk subjects aged 47.7 +/- 10.8 (range 30-65) years were performed before and after 4 min of upper arm occlusion. Concurrent with the ultrasound measurement (where color Doppler imaging was used to enhance arterial boundary detection), PPG signals were recorded from both index fingers for cross evaluation and comparison. Our results show that the finger PPG-AC exhibits a similar response to that of the well-known BA dilation: following the release of pressure (cuff around the BA), the PPG-AC increases abruptly before slowly decreasing toward the baseline. The peak PPG-AC is reached significantly faster than the peak FMD measured by ultrasound among healthy and risk groups (P < 0.001). The proposed technique using a finger photoplethysmogram can be applied in a rapid and non-invasive assessment of peripheral vascular functions as an alternative low-cost and less operator-dependent tool compared to ultrasound.
本研究旨在探讨手指光电容积脉搏波振幅(PPG-AC)与标准多普勒超声相比,在通过血流介导的血管舒张(FMD)评估内皮功能方面的实用性和有效性。对31名年龄在39.7±11.3岁(范围22 - 64岁)的健康受试者和52名年龄在47.7±10.8岁(范围30 - 65岁)的风险受试者的右肱动脉(BA)进行了高分辨率B模式扫描,扫描在肱动脉上臂阻断4分钟前后进行。在超声测量(使用彩色多普勒成像增强动脉边界检测)的同时,从两只食指记录PPG信号以进行交叉评估和比较。我们的结果表明,手指PPG-AC对著名的肱动脉扩张表现出类似的反应:在压力释放(肱动脉周围的袖带)后,PPG-AC会突然增加,然后缓慢下降至基线。在健康组和风险组中,PPG-AC峰值的达到速度明显快于超声测量的FMD峰值(P < 0.001)。所提出的使用手指光电容积脉搏波的技术可用于快速、无创地评估外周血管功能,作为一种与超声相比成本更低且对操作者依赖较小的替代工具。