使用自动验光仪和像差仪对早老性远视眼进行客观调节测量。

Objective accommodation measurements in prepresbyopic eyes using an autorefractor and an aberrometer.

作者信息

Win-Hall Dorothy M, Glasser Adrian

机构信息

University of Houston, College of Optometry, Houston, Texas, USA.

出版信息

J Cataract Refract Surg. 2008 May;34(5):774-84. doi: 10.1016/j.jcrs.2007.12.033.

Abstract

PURPOSE

To test the repeatability of the iTrace wavefront aberrometer (Tracey Technologies, Inc.) and the WR-5100K autorefractor (Grand Seiko Co., Ltd.) and to measure accomodation in young and phakic prepresbyopic subjects.

SETTING

University of Houston, College of Optometry, Houston, Texas, USA.

METHODS

This study comprised 30 young adults and 15 prepresbyopic subjects. Accommodation was stimulated with near charts presented at various distances. Measurements were repeated 3 times for each target distance with both instruments. For test-retest reliability, the entire protocol was repeated on 3 additional days in 3 prepresbyopic subjects as well as twice on the same day in the young adults.

RESULTS

The mean age was 25.5 years +/- 3.25 (SD) (range 21 to 31 years old) in the young adult group and 41.2 +/- 2.98 years (range 38 to 49 years) in the prepresbyopic group. Bland-Altman analysis of repeated measures of the young subjects had limits of agreements (LoA) of 1.58 diopters (D) or less for each instrument and when compared between instruments. Normalized mean stimulus response functions in the prepresbyopic group were similar for the 2 instruments. Bland-Altman analysis of the accommodation measurements between the 2 instruments showed a mean difference of 0.07 D and an LoA of 0.70 D. Repeated measures of 3 prepresbyopic subjects had a range of standard deviations from 0.07 to 0.51 D.

CONCLUSION

The accommodative responses measured with the 2 instruments were not significantly different, and testing showed both instruments to be suitable for objective measurement of accommodation in a phakic prepresbyopic population with low accommodative amplitudes.

摘要

目的

测试iTrace波前像差仪(Tracey Technologies公司)和WR - 5100K自动验光仪(精工光学株式会社)的可重复性,并测量年轻和有晶状体的早老性近视前期受试者的调节功能。

设置

美国得克萨斯州休斯顿大学视光学院。

方法

本研究包括30名年轻成年人和15名早老性近视前期受试者。通过在不同距离呈现近视力表来刺激调节。使用这两种仪器对每个目标距离的测量均重复3次。为了测试重测信度,在3名早老性近视前期受试者中,整个方案在另外3天重复进行,在年轻成年人中则在同一天重复进行2次。

结果

年轻成年人组的平均年龄为25.5岁±3.25(标准差)(范围为21至31岁),早老性近视前期组的平均年龄为41.2±2.98岁(范围为38至49岁)。对年轻受试者重复测量的Bland - Altman分析显示,每种仪器以及两种仪器之间比较时,一致性界限(LoA)均为1.58屈光度(D)或更低。早老性近视前期组中两种仪器的标准化平均刺激反应函数相似。两种仪器之间调节测量的Bland - Altman分析显示平均差异为0.07 D,LoA为0.70 D。3名早老性近视前期受试者的重复测量标准差范围为0.07至0.51 D。

结论

用这两种仪器测量的调节反应无显著差异,测试表明这两种仪器均适用于客观测量有晶状体的早老性近视前期人群低调节幅度的调节功能。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/b2e2/2917976/63e393366c4d/nihms220948f1.jpg

相似文献

3
10
Accommodation measurements in prepresbyopic and presbyopic populations.准老视和老视人群的调节测量
J Cataract Refract Surg. 2005 Sep;31(9):1682-3; author reply 1683. doi: 10.1016/j.jcrs.2005.09.004.

引用本文的文献

本文引用的文献

3
Stiffness gradient in the crystalline lens.晶状体中的硬度梯度。
Graefes Arch Clin Exp Ophthalmol. 2007 Sep;245(9):1357-66. doi: 10.1007/s00417-007-0537-1. Epub 2007 Feb 7.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验