Department of Endodontology, Oregon Health & Science University, School of Dentistry, 611 SW Campus Drive, Portland, OR 97239, USA.
J Endod. 2005 Jul;31(7):507-9. doi: 10.1097/01.don.0000152295.05827.80.
The purpose of this study was to compare the accuracy of the Root ZX and Elements Diagnostic electronic apex locators under clinical conditions. Thirty-six teeth planned for extraction were used. Each tooth was decoronated, coronally flared with Orifice Shapers, and irrigated with 2.6% sodium hypochlorite. Working lengths were measured with K-files using both electronic apex locators. The files were cemented at the last measured working length and the teeth were extracted. The apical 4-mm of each canal were exposed and photographed under 15x and 30x magnification. Images of each apex were projected and the distance from the file tip to the minor diameter was determined. The mean distances from the file tip to the minor diameter were 0.346 mm for the Elements Diagnostic and 0.410-mm for the Root ZX beyond the minor constriction. In locating the minor constriction the Root ZX was accurate 75% of the time to +/-0.5 mm, 83.3% +/-0.75 mm, and 88.9% to +/-1 mm. The Elements Diagnostic was accurate 75% of the time to +/-0.5 mm, 88.9% to +/-0.75 mm, and 91.7% to +/-1 mm. There was no statistically significant difference between the accuracy of the two electronic apex locators in locating the minor diameter (p < 0.05).
本研究的目的是比较Root ZX和Elements Diagnostic电子根尖定位仪在临床条件下的准确性。使用了36颗计划拔除的牙齿。每颗牙齿均进行去冠处理,用根管口扩大器进行冠部扩管,并使用2.6%的次氯酸钠冲洗。使用两种电子根尖定位仪,用K锉测量工作长度。将锉针粘固在最后测量的工作长度处,然后拔除牙齿。将每个根管的根尖4毫米暴露出来,在15倍和30倍放大倍数下拍照。投影每个根尖的图像,并确定锉针尖端到根管最小直径处的距离。在根管最小缩窄处之外,Elements Diagnostic的锉针尖端到根管最小直径处的平均距离为0.346毫米,Root ZX为0.410毫米。在定位根管最小缩窄处时,Root ZX在±0.5毫米范围内的定位准确率为75%,在±0.75毫米范围内为83.3%,在±1毫米范围内为88.9%。Elements Diagnostic在±0.5毫米范围内的定位准确率为75%,在±0.75毫米范围内为88.9%,在±1毫米范围内为91.7%。两种电子根尖定位仪在定位根管最小直径处的准确性之间无统计学显著差异(p<0.05)。