An in vitro comparison of root canal length determination by DentaPort ZX and iPex apex locators.

作者信息

Puri Nikhil, Chadha Rupali, Kumar Pragya, Puri Komal

机构信息

Department of Conservative Dentistry and Endodontics, Institute of Dental Studies and Technologies, Modinagar, Uttar Pradesh, India.

Consultant Endodontist, New Delhi, India.

出版信息

J Conserv Dent. 2013 Nov;16(6):555-8. doi: 10.4103/0972-0707.120953.

Abstract

AIM

THE AIM OF THE PRESENT STUDY IS TO EVALUATE AND TO COMPARE THE ACCURACY OF TWO ELECTRONIC APEX LOCATORS (EALS): DentaPort ZX and iPex, at a position 0.5 mm short of the apical foramen.

MATERIALS AND METHODS

Thirty single-rooted, mandibular premolar teeth were selected. Standard access cavities were prepared and the actual length (AL) was calculated. The samples were then embedded in alginate and the electronic measurements were determined and recorded.

RESULTS

The results obtained showed that in determining the root canal length with a tolerance level of ±0.5 mm, i.e., AL ± 0.5 mm, DentaPort ZX was accurate in 93.3% of the samples and iPex was accurate in 90% of the samples at a position 0.5 mm short of the apical foramen.

CONCLUSION

A strong correlation was seen between the two electronic methods and AL and also in between the two EALs, showing the possibility of their use to measure the root canal length. No statistically significant difference was found between both the apex locators.

摘要

目的

本研究的目的是评估和比较两种电子根尖定位仪(EALS):DentaPort ZX和iPex在距离根尖孔0.5毫米处的准确性。

材料与方法

选择30颗单根下颌前磨牙。制备标准的开髓洞形并计算实际长度(AL)。然后将样本嵌入藻酸盐中,进行电子测量并记录。

结果

所得结果表明,在以±0.5毫米的公差水平确定根管长度,即AL±0.5毫米时,DentaPort ZX在距离根尖孔0.5毫米处对93.3%的样本测量准确,iPex对90%的样本测量准确。

结论

两种电子方法与AL之间以及两种EAL之间均存在强相关性,表明它们可用于测量根管长度。两种根尖定位仪之间未发现统计学上的显著差异。

相似文献

1
An in vitro comparison of root canal length determination by DentaPort ZX and iPex apex locators.
J Conserv Dent. 2013 Nov;16(6):555-8. doi: 10.4103/0972-0707.120953.
2
The Effect of Long Service Life on the Accuracy of Dentaport ZX Electronic Apex Locator.
Iran Endod J. 2022;17(4):195-199. doi: 10.22037/iej.v17i4.34623.
3
Evaluation of the accuracy of different apex locators in determining the working length during root canal retreatment.
J Dent Res Dent Clin Dent Prospects. 2020 Spring;14(2):125-129. doi: 10.34172/joddd.2020.026. Epub 2020 Jun 17.
4
In vivo evaluation of the iPex and Root ZX electronic apex locators using various irrigants.
Int Endod J. 2013 Aug;46(8):769-74. doi: 10.1111/iej.12057. Epub 2013 Mar 28.
8
Evaluation of Dentaport ZX and Raypex 6 electronic apex locators: an in vivo study.
Med Oral Patol Oral Cir Bucal. 2014 Mar 1;19(2):e202-5. doi: 10.4317/medoral.19114.
9
A comparative assessment of the accuracies of Raypex 5, Raypex 6, iPex and iPex II electronic apex locators: An in vitro study.
J Istanb Univ Fac Dent. 2017 Jan 2;51(1):28-33. doi: 10.17096/jiufd.61309. eCollection 2017.
10
Accuracy of a new apex locator: an in vitro study.
Int Endod J. 2002 Feb;35(2):186-92. doi: 10.1046/j.1365-2591.2002.00468.x.

引用本文的文献

2
Accuracy of different generations of apex locators in determining working length; a systematic review and meta-analysis.
Saudi Dent J. 2022 Jan;34(1):11-20. doi: 10.1016/j.sdentj.2021.09.020. Epub 2021 Sep 25.
3
Evaluation of the accuracy of different apex locators in determining the working length during root canal retreatment.
J Dent Res Dent Clin Dent Prospects. 2020 Spring;14(2):125-129. doi: 10.34172/joddd.2020.026. Epub 2020 Jun 17.
5
A comparative assessment of the accuracies of Raypex 5, Raypex 6, iPex and iPex II electronic apex locators: An in vitro study.
J Istanb Univ Fac Dent. 2017 Jan 2;51(1):28-33. doi: 10.17096/jiufd.61309. eCollection 2017.
6
Influence of root canal curvature on the accuracy of an electronic apex locator: An in vitro study.
J Conserv Dent. 2014 Nov;17(6):583-6. doi: 10.4103/0972-0707.144610.

本文引用的文献

1
Exact Method of Locating the Apical Foramen.
Dent Regist. 1918 Sep;72(9):420-425.
2
Efficiency of 2 electronic apex locators on working length determination: A clinical study.
J Conserv Dent. 2013 May;16(3):229-32. doi: 10.4103/0972-0707.111320.
4
An evaluation of root ZX and iPex apex locators: an in vivo study.
J Endod. 2011 May;37(5):608-10. doi: 10.1016/j.joen.2011.02.002. Epub 2011 Mar 22.
6
An ex vivo comparison of root canal length determination by three electronic apex locators at positions short of the apical foramen.
Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2010 Aug;110(2):e57-61. doi: 10.1016/j.tripleo.2010.03.005. Epub 2010 Jun 23.
7
In vitro evaluation of the accuracy of two electronic apex locators.
J Conserv Dent. 2008 Apr;11(2):82-5. doi: 10.4103/0972-0707.44056.
8
An in vivo comparison of working length determination of two frequency-based electronic apex locators.
Int Endod J. 2009 Nov;42(11):1026-31. doi: 10.1111/j.1365-2591.2009.01615.x.
10
Outcome of primary root canal treatment: systematic review of the literature -- Part 2. Influence of clinical factors.
Int Endod J. 2008 Jan;41(1):6-31. doi: 10.1111/j.1365-2591.2007.01323.x. Epub 2007 Oct 10.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验