Brockmann Claudia, Schulz Marcus, Laube Thomas
Department of Ophthalmology, Duisburg-Essen University, Essen, and Varian, Inc., Darmstadt, Germany.
J Cataract Refract Surg. 2008 Jul;34(7):1161-6. doi: 10.1016/j.jcrs.2008.03.039.
To record the spectral transmittance curves of various ultraviolet (UV) and blue-light-filtering intraocular lenses (IOLs), evaluate their UV and blue-light-radiation- absorption capacities, and compare them with those of the natural crystalline lens.
Department of Ophthalmology, Duisburg-Essen University, Essen, Germany.
Eight IOLs from 6 manufacturers were evaluated. The transmittance of wavelengths from 200 nm to 800 nm was measured using a high-performance spectrophotometer with a diffuse transmittance accessory through a 1.5 mm aperture.
All assessed IOLs provided good UVC (200 to 280 nm) and UVB (280 to 320 nm) protection, but 2 IOLs lacked sufficient UVA (320 to 400 nm) protection. Major differences in the absorption capacities were observed in the blue-light range (400 to 500 nm).
The absorption characteristics of some UV and blue-light-filtering IOLs resembled those of the crystalline lens, but some differed. Long-term clinical trials should be performed to determine how blue-blocking IOLs affect the risk for progression of age-related macular degeneration.
记录各种紫外线(UV)和蓝光滤过型人工晶状体(IOL)的光谱透过率曲线,评估其紫外线和蓝光辐射吸收能力,并与天然晶状体的吸收能力进行比较。
德国埃森杜伊斯堡 - 埃森大学眼科。
对来自6家制造商的8种人工晶状体进行评估。使用配备漫射透过率附件的高性能分光光度计,通过1.5毫米孔径测量200纳米至800纳米波长的透过率。
所有评估的人工晶状体都提供了良好的UVC(200至280纳米)和UVB(280至320纳米)防护,但有2种人工晶状体缺乏足够的UVA(320至400纳米)防护。在蓝光范围(400至500纳米)观察到吸收能力存在主要差异。
一些紫外线和蓝光滤过型人工晶状体的吸收特性与晶状体相似,但也有一些不同。应进行长期临床试验,以确定蓝光阻断型人工晶状体如何影响年龄相关性黄斑变性进展的风险。