LaFrance Martin W, Kehinde Lucy E, Fullard Roderick J
School of Optometry, University of Alabama at Birmingham, Birmingham, Alabama, USA.
Curr Eye Res. 2008 Jul;33(7):525-44. doi: 10.1080/02713680802190085.
To compare instruments and reagents used for cytometric bead-based assay (CBA) of cytokines in human tears.
Two CBA instruments and reagent kits from five vendors were compared in parallel. Vendor protocols were modified to accommodate tear volumes. Precision, recovery, dilution linearity, and stability were evaluated.
Luminextrade mark and Bio-Radtrade mark proved optimal for tear CBA. Good correlation for dilution linearity was observed (r > 0.80, p < 0.05) for 1-4-mu l sample volumes. Samples were stable for 30 min at 25 degrees C.
Many tear cytokines can be reliably measured by CBA using an optimized protocol. Concentrations remain stable during periods consistent with collection.