Lohmann A, Diekmann N, Walters R, Kraft R
R & D, Referat Pharmakokinetik, Thiemann Arzneimittel GmbH, Waltrop, Germany.
J Chromatogr. 1991 Mar 8;564(1):283-8. doi: 10.1016/0378-4347(91)80092-q.
A sensitive and specific method for the determination of bencyclane in human plasma is presented. Bencyclane was extracted from human plasma with two 3-ml volumes of isooctane and was shaken for 10 min. The organic phase was separated and evaporated to dryness at 40 degrees C under a nitrogen stream. The residue was dissolved and an aliquot was injected into the gas chromatograph. The separation was performed with a DB-17 column with helium as the carrier gas. Nitrogen-selective detection was performed. The quantification was performed with the signal output. The limit of detection was 1 ng/ml.
本文介绍了一种灵敏且特异的测定人血浆中苄环烷的方法。用两体积3毫升的异辛烷从人血浆中提取苄环烷,并振荡10分钟。分离有机相,在氮气流下于40℃蒸发至干。将残渣溶解,取一份注入气相色谱仪。用DB-17柱进行分离,以氦气作为载气。进行氮选择性检测。通过信号输出进行定量。检测限为1纳克/毫升。