Alkurt Meryem Toraman, Peker Ilkay, Usalan Gulten, Altunkaynak Bulent
Department of Oral Diagnosis and Radiology, Faculty of Dentistry, Gazi University, Ankara, Turkey.
J Contemp Dent Pract. 2008 Jul 1;9(5):34-41.
The purpose of this study was to evaluate the effect of tube current reduction on image quality using medium and regular intensifying screens as well as a digital system for panoramic radiography.
A total of 150 panoramic images of 75 patients were obtained in the study. The initial images were taken at standard exposure settings, and secondary images were exposed with the tube current reduced at different rates.
There was no statistically significant difference (p>0.05) between the two exposures for Group 3 (the rate of dose reduction 25%) while a statistically significant difference (p<0.05) was found in Group 4 (the rate of dose reduction 50%) using medium intensifying screens for all observers. No statistically significant difference was found between the two exposures on digital panoramic images.
According to the results of this study a dose reduction of 25% was achieved for medium intensifying screens and for digital panoramic images without any loss of image quality.
A substantial reduction in radiation exposure can be achieved in conventional panoramic radiography using a medium intensifying screen and in digital panoramic radiography without any loss of image quality needed for radiological evaluation of anatomical structures and pathological conditions.
本研究旨在评估使用中速和常规增感屏以及全景数字化摄影系统降低管电流对图像质量的影响。
本研究共获取了75例患者的150张全景图像。初始图像在标准曝光设置下拍摄,二次图像以不同速率降低管电流进行曝光。
对于使用中速增感屏的所有观察者,第3组(剂量降低率25%)的两次曝光之间无统计学显著差异(p>0.05),而第4组(剂量降低率50%)有统计学显著差异(p<0.05)。数字化全景图像的两次曝光之间未发现统计学显著差异。
根据本研究结果,对于中速增感屏和数字化全景图像,剂量降低25%时图像质量无任何损失。
在传统全景摄影中使用中速增感屏以及在数字化全景摄影中,可大幅降低辐射剂量,而不会损失对解剖结构和病理状况进行放射学评估所需的图像质量。