Predictive, error-compensating kinetic method for enzymatic quantification of creatinine in serum.

作者信息

Bacon B L, Pardue H L

机构信息

Department of Chemistry, Purdue University, West Lafayette, IN 47907-1393.

出版信息

Clin Chem. 1991 Aug;37(8):1338-44.

PMID:1868593
Abstract

Here we describe an error-compensating kinetic-based method for the enzymatic quantification of creatinine in serum. The method, which has a large linear range and very low dependency on experimental variables that influence enzyme activity, is based on the use of creatinine amidohydrolase in a four-step coupled reaction sequence to generate a product that is monitored photometrically. We collected data for absorbance vs time during two to four half-lives of each reaction and fit a first-order model to the data to compute the total absorbance change that would be measured if the reaction were monitored to completion. Computed values of absorbance change agreed well with measured values and varied linearly with creatinine concentration in the sample throughout the range examined: 44 to 1326 mumol/L. A twofold change in enzyme activity in the final reaction mixture (3.3 to 6.6 kU/L) produced changes of only 12% and 4% for creatinine concentrations of 44 and 353 mumol/L, respectively. Results (y) for 39 serum samples that contained creatinine concentrations between 20 and 1800 mumol/L agreed well with liquid-chromatographic results (x), yielding linear least-squares statistics of y = (1.03 +/- 0.01) x + (5 +/- 5) mumol/L (r = 0.995, Sy.x = 37 mumol/L). We conclude that the predictive kinetic approach is a robust method for the quantification of creatinine in serum.

摘要

相似文献

1
Predictive, error-compensating kinetic method for enzymatic quantification of creatinine in serum.
Clin Chem. 1991 Aug;37(8):1338-44.
2
A nonlinear regression-kinetic method for quantification of serum triglycerides.
Clin Chem. 1983 Jul;29(7):1392-5.
3
Quantitation of lactate by a kinetic method with an extended range of linearity and low dependence on experimental variables.采用动力学方法对乳酸进行定量分析,该方法具有更宽的线性范围且对实验变量依赖性低。
Clin Chem. 1984 Feb;30(2):226-9.
4
An enzymic, reaction-rate assay for serum creatinine with a centrifugal analyzer.一种使用离心分析仪的血清肌酐酶促反应速率测定法。
Clin Chem. 1982 Jan;28(1):114-7.
5
Determination of creatinine in serum and urine by cation-exchange high-pressure liquid chromatography.采用阳离子交换高压液相色谱法测定血清和尿液中的肌酐。
Clin Chem. 1991 Apr;37(4):563-5.
6
Evaluation of a predictive kinetic procedure for the enzymatic measurement of creatinine in serum.血清中肌酐酶法测定的预测动力学方法评估
Clin Chem Lab Med. 1998 Oct;36(10):781-8. doi: 10.1515/CCLM.1998.139.
7
Enzymatic correction of interference in the kinetic Jaffé reaction for determining creatinine in plasma.用于测定血浆中肌酐的动力学Jaffé反应中干扰的酶法校正
Clin Chem. 1985 Sep;31(9):1564-5.
8
Determination of creatinine in serum and urine by a rapid liquid-chromatographic method.
Clin Chem. 1990 Jun;36(6):830-6.
9
High-performance liquid chromatographic determination of creatinine.高效液相色谱法测定肌酐
Scand J Clin Lab Invest. 1991 Feb;51(1):67-71.
10
Multicentre evaluation of an enzymatic method for creatinine determination using a sensitive colour reagent.使用灵敏显色试剂的肌酐测定酶法的多中心评估
J Clin Chem Clin Biochem. 1986 Nov;24(11):889-902.