Probe electrospray ionization (PESI) has recently been developed, in which the electrospray was generated from a solid needle instead of by using a capillary. In this paper, the characteristics of probe electrospray ionization were studied based on the measurement of spray current, optical microscopy, and PESI mass spectrometry. In the experiment, the solid needle was moved up and down a vertical axis, and a small amount of sample was repeatedly loaded to the needle when the tip of the needle touched the surface of the liquid sample at the lowest position. After the application of high voltage, a liquid droplet was formed on the tip of the solid needle probe, with its size was determined by the size of the needle tip. The liquid flow rate to the tip, as indicated by the spray current, depends on the voltage applied to the needle as well as the loaded liquid amount. Stable electrospray can be maintained until the total consumption of liquid sample. The kilohertz current pulsation takes place in the case of overloading the sample to the needle. The influences of the applied voltage and the liquid flow rate on the PESI mass spectra were also examined.
探针电喷雾电离(PESI)是最近开发出来的,其中电喷雾是由实心针产生的,而不是使用毛细管。本文基于喷雾电流测量、光学显微镜和PESI质谱对探针电喷雾电离的特性进行了研究。在实验中,实心针在垂直轴上上下移动,当针的尖端在最低位置接触液体样品表面时,将少量样品反复加载到针上。施加高压后,在实心针探针的尖端形成一个液滴,其大小由针尖的大小决定。如喷雾电流所示,流向尖端的液体流速取决于施加在针上的电压以及加载的液体量。在液体样品完全消耗之前,可以维持稳定的电喷雾。在样品过载到针的情况下会发生千赫兹电流脉动。还研究了施加电压和液体流速对PESI质谱的影响。