Anderson M S, Swenson C A
Ames Laboratory-USDOE and Department of Physics, Iowa State University, Ames, Iowa 50011.
Rev Sci Instrum. 1978 Aug;49(8):1027. doi: 10.1063/1.1135517.
A low-temperature comparator cryostat with a self-balancing ac resistance bridge and a data acquisition system have been used over the past several years to calibrate approximately 45 germanium resistance thermometers, many of them several times. Examples are given of typical thermometer reproducibility results, together with algorithms for estimating self-heating and ac-dc resistance differences for these thermometers. Recent experiences with rhodium-iron resistance thermometry are described in terms of a comparison of previous ISU magnetic thermometry with gas thermometry results from the National Physical Laboratory in England.
在过去几年中,一直使用带有自平衡交流电阻桥和数据采集系统的低温比较式低温恒温器来校准大约45支锗电阻温度计,其中许多温度计都校准了好几次。文中给出了典型温度计再现性结果的示例,以及估算这些温度计自热和交直流电阻差异的算法。根据之前爱荷华州立大学的磁温度计测量结果与英国国家物理实验室的气体温度计测量结果的比较,描述了铑铁电阻温度计的最新经验。