Accuracy of colour Doppler ultrasound velocity measurements in small vessels.

作者信息

Deane C R, Forsberg F, Thomas N, Roberts V C

机构信息

Department of Medical Physics and Medical Engineering, King's College School of Medicine and Dentistry, Dulwich Hospital, London, UK.

出版信息

J Biomed Eng. 1991 May;13(3):249-54. doi: 10.1016/0141-5425(91)90136-u.

Abstract

Colour Doppler ultrasound offers the possibility of imaging small vessels not visible by B-mode alone. The colour Doppler image of velocities allows the course of small vessels to be imaged in the X-Y plane of the scan provided the Doppler frequency shift is of sufficient magnitude. This permits alignments of the Doppler cursor, allowing angle correction to provide true velocity measurements from the Doppler shift obtained. Before attempting to make velocity measurements, however, it is essential to be aware of the possible error in the Z plane caused by the thickness of the Doppler sample volume. To quantify this source of error, hydrophone and flow-rig measurements were performed on an Acuson 128 colour Doppler scanner with both 5 MHz linear-array and 3.5 MHz phased-array transducers. Measurements of the transmitted pulses using a point hydrophone showed that both probes employ approximately 3.5 MHz Doppler pulses (in both colour and pulsed Doppler modes). The two transducers have the same axial resolution. In colour Doppler mode the axial length of the sample volume increases automatically with depth by up to 0.5 mm. Measurements of colour and pulsed Doppler signal strength were obtained in a controlled flow rig. Both transducers produced accurate colour flow images of the phantom at their optimum depths; flow velocity errors due to Z-plane thickness are less than 5%. There was, however, substantial error outside these optimum conditions (up to 20%).

摘要

彩色多普勒超声提供了对仅用B模式无法看到的小血管进行成像的可能性。速度的彩色多普勒图像使得只要多普勒频移幅度足够大,就能在扫描的X-Y平面上对小血管的走行进行成像。这允许对齐多普勒光标,从而进行角度校正,以便根据获得的多普勒频移提供真实的速度测量值。然而,在尝试进行速度测量之前,必须意识到由于多普勒采样容积厚度在Z平面上可能产生的误差。为了量化这种误差来源,使用水听器和流量装置在配备5 MHz线性阵列和3.5 MHz相控阵换能器的Acuson 128彩色多普勒扫描仪上进行了测量。使用点水听器对发射脉冲进行的测量表明,两种探头均采用约3.5 MHz的多普勒脉冲(彩色和脉冲多普勒模式下均如此)。这两种换能器具有相同的轴向分辨率。在彩色多普勒模式下,采样容积的轴向长度会随着深度自动增加,最大可达0.5 mm。在可控流量装置中获得了彩色和脉冲多普勒信号强度的测量值。两种换能器在其最佳深度处均能产生准确的体模彩色血流图像;由于Z平面厚度导致的流速误差小于5%。然而,在这些最佳条件之外存在相当大的误差(高达20%)。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验