Abstract

Pulsed wave (PW) Doppler ultrasound is routinely used in the clinic to assess blood flow. Our annual Doppler quality assurance tests revealed unexpectedly large errors in measurement of maximum velocity, exceeding our tolerance (error >20%), when using certain scanners with small Doppler sample volume dimensions. The aim of this study was to assess the dependence of maximum velocity estimates on PW Doppler sample volume size. A flow phantom with known steady flow was used to acquire maximum velocity estimates (maximum velocities of 24, 39 and 85 cm/s and sample volume range of 0.3-20 mm) with a variety of transducers and scanners in clinical use (51 probes from 4 manufacturers). Selected acoustic outputs were characterized using free-field hydrophone measurements. All maximum velocity estimates were within our tolerance for sample volume sizes ≥1.5 mm, although maximum velocity estimates typically increased with decreasing sample volume size. Errors exceeding our tolerance were commonly found for one manufacturer when using smaller sample volumes, resulting in up to 75% overestimation. Although intrinsic spectral broadening based on transit time considerations may help explain our findings, the sample volume dependence raises potential clinical concerns that users should be aware of and which manufacturers should consider addressing.

摘要

脉冲波(PW)多普勒超声在临床上常用于评估血流。我们的年度多普勒质量保证测试显示,当使用某些具有较小多普勒取样体积尺寸的扫描仪时,测量最大速度的测量误差大得惊人,超出了我们的容忍度(误差>20%)。本研究旨在评估最大速度估计值对 PW 多普勒取样体积大小的依赖性。使用具有已知稳定流量的流量模型,使用各种临床使用中的探头和扫描仪(来自 4 个制造商的 51 个探头)获取最大速度估计值(最大速度为 24、39 和 85cm/s,取样体积范围为 0.3-20mm)。使用自由场水听器测量来对选定的声输出进行特性描述。尽管最大速度估计值通常随取样体积尺寸的减小而增加,但对于≥1.5mm 的取样体积尺寸,所有最大速度估计值都在我们的容忍度范围内。当使用较小的取样体积时,一个制造商的误差通常超出了我们的容忍度,最大误差高达 75%。尽管基于渡越时间考虑的固有频谱展宽可能有助于解释我们的发现,但取样体积的依赖性引起了潜在的临床关注,用户应该意识到这一点,制造商也应该考虑解决这一问题。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验