Razeghinejad Mohammad Reza, Salouti Ramin, Khalili Mohammad Reza
Poostchi Ophthalmology Research Center, Shiraz University of Medical Sciences, Shiraz, Iran.
Iran J Med Sci. 2014 Jan;39(1):11-9.
To evaluate the agreement in intraocular pressure (IOP) measurements by Ocular Response Analyzer (ORA) and Tono-Pen XL (TXL) with the Goldmann Applanation Tonometer (GAT) and to examine corneal biomechanical properties in aphakic glaucoma patients with a central corneal thickness (CCT) >600 µ.
Thirty-six eyes of aphakic glaucoma patients (group 1) and 40 eyes of normal children (group 2) were studied. The mean ORA and TXL IOP values were compared with the GAT-IOP values. Regression analyses were used to evaluate the associations between IOP and CCT, corneal hysteresis (CH), and corneal resistance factor (CRF). Bland-Altman plots were used to evaluate the agreement between the tonometers.
The mean±standard deviations of the age and male/female ratio were 16.58±5.44 and 15.75±5.04 years and 14/22 and 18/22 in group 1 and group 2, respectively. CCT in group 1 was 651.1±42 and in group 2 was 567.3±32.4. In group 1, the mean TXL (22.4, P=0.004), IOPcc (corneal compensated) (27.8, P=0.005), and IOPg (Goldmann correlated) values (28.1, P<0.0001) were greater than GAT-IOP (20.6). In group 2, only IOPg value (16.4) was higher than GAT-IOP (14.8, P=0.04). IOP reading of all the tonometers were positively and negatively associated with CRF and CH in the multiple regression analysis, respectively.
The TXL had a greater agreement with the GAT, and the ORA overestimated IOP in aphakic glaucoma patients. The ORA and TXL seemed to be affected by CH and CRF.
评估眼反应分析仪(ORA)和非接触眼压计XL(TXL)测量眼压(IOP)与Goldmann压平眼压计(GAT)测量结果的一致性,并研究中央角膜厚度(CCT)>600µ的无晶状体性青光眼患者的角膜生物力学特性。
研究了36例无晶状体性青光眼患者的36只眼(第1组)和40例正常儿童的40只眼(第2组)。将ORA和TXL的平均IOP值与GAT-IOP值进行比较。采用回归分析评估IOP与CCT、角膜滞后(CH)和角膜阻力因子(CRF)之间的关联。采用Bland-Altman图评估眼压计之间的一致性。
第1组和第2组的年龄均值±标准差分别为16.58±5.44岁和15.75±5.04岁,男女比例分别为14/22和18/2;第1组的CCT为651.1±42,第2组为567.3±32.4。在第1组中,TXL的平均眼压值(22.4,P=0.004)、角膜补偿眼压值(IOPcc,27.8,P=0.005)和Goldmann相关眼压值(IOPg,28.1,P<0.0001)均高于GAT-IOP(20.6)。在第2组中,仅IOPg值(16.4)高于GAT-IOP(14.8,P=0.04)。在多元回归分析中,所有眼压计的IOP读数分别与CRF呈正相关,与CH呈负相关。
TXL与GAT的一致性更好,而ORA高估了无晶状体性青光眼患者的眼压。ORA和TXL似乎受CH和CRF的影响。