健康受试者动脉血二氧化碳分压急性变化后经皮二氧化碳反应的评估。

Evaluation of transcutaneous CO2 responses following acute changes in PaCO2 in healthy subjects.

作者信息

Fuke Satoshi, Miyamoto Kenji, Ohira Hiroshi, Ohira Megumi, Odajima Nao, Nishimura Masaharu

机构信息

First Department of Medicine, Hokkaido University School of Medicine, Sapporo, Japan.

出版信息

Respirology. 2009 Apr;14(3):436-42. doi: 10.1111/j.1440-1843.2008.01442.x. Epub 2009 Dec 16.

Abstract

BACKGROUND AND OBJECTIVE

Transcutaneous blood gas-monitoring systems with miniaturized SpO(2) (peripheral blood oxygen saturation)/PCO(2) combined sensors (TOSCATM) have been widely used. There are no reports of the inter- and intra-individual variability in transcutaneous measurements of PaCO(2) (PtcCO(2)) in response to acute progressive changes in PaCO(2). This study examined inter- and intra-individual variability of PtcCO(2) measurements under semi-steady-state conditions, and characterized the behaviour of PtcCO(2) in response to acute progressive changes in PaCO(2).

METHODS

Subjects breathed mixed gases through a mouthpiece connected to an automatic arterial blood gas controller. Using end-tidal PCO(2) and PO(2) as guides, PaCO(2) was controlled to increase and/or decrease between baseline and > or = 60 mm Hg, in a stepwise (n = 9) or progressive fashion (n = 6). Arterial blood was sampled when needed.

RESULTS

Intra-individual correlation coefficients between PtcCO(2) and PaCO(2) were excellent in all subjects (0.971-0.989); however, the slope of the regression line varied among subjects (1.040-1.335). Bias and limits of agreement (+/- 2 SD from bias) between PtcCO(2) and PaCO(2) were -1.8 mm Hg and -7.7 to 4.1 mm Hg. Changes in PtcCO(2) in response to acute progressive changes in PaCO(2) also varied among subjects.

CONCLUSION

The PtcCO(2) measurement system allows reliable estimation of PaCO(2) in a given subject. However, caution is needed when comparing absolute values between subjects or when acute changes in PaCO(2) occur.

摘要

背景与目的

带有小型化SpO₂(外周血氧饱和度)/PCO₂组合传感器的经皮血气监测系统(TOSCATM)已被广泛应用。目前尚无关于经皮测量PaCO₂(PtcCO₂)以响应PaCO₂急性渐进性变化时个体间和个体内变异性的报道。本研究检测了半稳态条件下PtcCO₂测量的个体间和个体内变异性,并描述了PtcCO₂响应PaCO₂急性渐进性变化的行为。

方法

受试者通过连接到自动动脉血气控制器的咬嘴呼吸混合气体。以呼气末PCO₂和PO₂为指导,将PaCO₂控制在基线至≥60 mmHg之间逐步(n = 9)或渐进性(n = 6)增加和/或降低。必要时采集动脉血样本。

结果

所有受试者中PtcCO₂与PaCO₂之间的个体内相关系数均极佳(0.971 - 0.989);然而,回归线的斜率在受试者之间有所不同(1.040 - 1.335)。PtcCO₂与PaCO₂之间的偏差及一致性界限(偏差±2 SD)为 -1.8 mmHg和 -7.7至4.1 mmHg。PtcCO₂响应PaCO₂急性渐进性变化的情况在受试者之间也存在差异。

结论

PtcCO₂测量系统能够可靠地估计特定受试者的PaCO₂。然而,在比较受试者之间的绝对值或PaCO₂发生急性变化时需要谨慎。