Abstract

Fiber optic catheters for the diagnosis of gastrointestinal motility disorders are demonstrated in-vitro and in-vivo. Single element catheters have been verified against existing solid state catheters and a multi-element catheter has been demonstrated for localized and full esophageal monitoring. The multi-element catheter consists of a series of closely spaced pressure sensors that pick up the peristaltic wave traveling along the gastrointestinal (GI) tract. The sensors are spaced on a 10 mm pitch allowing a full interpolated image of intraluminal pressure to be generated. Details are given of in-vivo trials of a 32-element catheter in the human oesophagus and the suitability of similar catheters for clinical evaluation in other regions of the human digestive tract is discussed. The fiber optic catheter is significantly smaller and more flexible than similar commercially available devices making intubation easier and improving patient tolerance during diagnostic procedures.

摘要

用于诊断胃肠动力障碍的光纤导管已在体外和体内得到验证。单元件导管已与现有的固态导管进行了对比验证,并且已展示了一种用于局部和全食管监测的多元件导管。多元件导管由一系列紧密间隔的压力传感器组成,这些传感器可检测沿胃肠道(GI)传播的蠕动波。传感器的间距为10毫米,可生成腔内压力的完整插值图像。文中给出了一种32元件导管在人体食管中的体内试验细节,并讨论了类似导管在人体消化道其他区域进行临床评估的适用性。光纤导管比类似的市售设备明显更小、更灵活,使插管更容易,并提高了诊断过程中患者的耐受性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验