Wali N, Leguire L E
Department of Ophthalmology, Children's Hospital, Columbus, OH 43205.
Doc Ophthalmol. 1991;76(4):367-75. doi: 10.1007/BF00142675.
Normative dark-adapted electroretinograms were recorded simultaneously with a skin electrode and corneal electrode for varying stimulus intensities. The electroretinogram b-wave amplitudes for each electrode were fitted by the Naka-Rushton equation, and the parameters Vmax, K and n were evaluated. A comparison of parameters between the two electrodes showed a significant difference for Vmax and K but not for n. Vmax was approximately eight times smaller and K was 0.3 log unit smaller for the skin electrode than for the corneal electrode. B-wave amplitude and implicit time were also compared between the two electrodes. The b-wave amplitude ratio of the corneal electrode to that of the skin electrode increased with luminance and ranged from 1.83 to 7.68. Overall, b-wave implicit time for the skin electrode was approximately 10 ms shorter than that of the corneal electrode.