巴西里约热内卢乳腺X线摄影检查中入射表面空气比释动能的评估。

Evaluation of the entrance surface air kerma in mammographic examinations in Rio de Janeiro, Brazil.

作者信息

Oliveira L C G, Dias T K, Lopes R T, Kodlulovich S

机构信息

Laboratório de Instrumentação Nuclear (LIN/COPPE/UFRJ), Centro de Tecnologia, Bloco I, Sala 133 Cidade Universitária, Ilha do Fundão, Rio de Janeiro, CEP 21949-900 RJ, Brasil.

出版信息

Radiat Prot Dosimetry. 2009 Feb;133(3):136-43. doi: 10.1093/rpd/ncp031. Epub 2009 Mar 14.

Abstract

The aim of this work was to evaluate the distribution of the entrance surface air kerma (ESAK) and the average glandular dose (D(G)) in four mammography facilities located in the city of Rio de Janeiro. The ESAK values were estimated from the X-ray tube output rate (mGy/mAs) parameters. The image quality was evaluated by the radiologists in each clinic. The ESAK values obtained for a breast thickness of 45 mm were 5.58 mGy in Clinic A, 10.07 mGy in Clinic B, 13.89 mGy in Clinic C and 7.21 mGy in Clinic D. For D(G), it can be seen that, for the same compressed breast thickness (50 mm), the value varied from 0.20 to 3.60 mGy, with a mean value of 1.50 mGy for all the clinics. In image quality evaluation, Clinic D was the only one that presented a very low acceptability for quality criteria and inadequacies in relation to specks, masses and optical density.

摘要

这项工作的目的是评估位于里约热内卢市的四家乳腺摄影机构的入射表面空气比释动能(ESAK)和平均腺体剂量(D(G))的分布情况。ESAK值是根据X射线管输出率(mGy/mAs)参数估算得出的。各诊所的放射科医生对图像质量进行了评估。对于45毫米乳房厚度所获得的ESAK值,诊所A为5.58 mGy,诊所B为10.07 mGy,诊所C为13.89 mGy,诊所D为7.21 mGy。对于D(G),可以看出,对于相同的压缩乳房厚度(50毫米),该值在0.20至3.60 mGy之间变化,所有诊所的平均值为1.50 mGy。在图像质量评估中,诊所D是唯一一家在质量标准方面可接受性极低且在斑点、肿块和光学密度方面存在不足的机构。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验