Das Mini, Gifford Howard C, O'Connor J Michael, Glick Stephen J
Department of Radiology, University of Massachusetts Medical School, 55 Lake Avenue North, Worcester, Massachusetts 01655, USA.
Med Phys. 2009 Jun;36(6):1976-84. doi: 10.1118/1.3116902.
In this article the authors evaluate a recently proposed variable dose (VD)-digital breast tomosynthesis (DBT) acquisition technique in terms of the detection accuracy for breast masses and microcalcification (MC) clusters. With this technique, approximately half of the total dose is used for one center projection and the remaining dose is split among the other tomosynthesis projection views. This acquisition method would yield both a projection view and a reconstruction view. One of the aims of this study was to evaluate whether the center projection alone of the VD acquisition can provide equal or superior MC detection in comparison to the 3D images from uniform dose (UD)-DBT. Another aim was to compare the mass-detection capabilities of 3D reconstructions from VD-DBT and UD-DBT. In a localization receiver operating characteristic (LROC) observer study of MC detection, the authors compared the center projection of a VD acquisitioh scheme (at 2 mGy dose) with detector pixel size of 100 microm with the UD-DBT reconstruction (at 4 mGy dose) obtained with a voxel size of 100 microm. MCs with sizes of 150 and 180 microm were used in the study, with each cluster consisting of seven MCs distributed randomly within a small volume. Reconstructed images in UD-DBT were obtained from a projection set that had a total of 4 mGy dose. The current study shows that for MC detection, using the center projection alone of VD acquisition scheme performs worse with area under the LROC curve (AL) of 0.76 than when using the 3D reconstructed image using the UD acquisition scheme (AL=0.84). A 2D ANOVA found a statistically significant difference (p=0.038) at a significance level of 0.05. In the current study, although a reconstructed image was also available using the VD acquisition scheme, it was not used to assist the MC detection task which was done using the center projection alone. In the case of evaluation of detection accuracy of masses, the reconstruction with VD-DBT (AL=0.71) was compared to that obtained from the UD-DBT (AL=0.78). The authors found no statistically significant difference between the two (p-value=0.22), although all the observers performed better for UD-DBT.
在本文中,作者评估了一种最近提出的可变剂量(VD)-数字乳腺断层合成(DBT)采集技术在检测乳腺肿块和微钙化(MC)簇方面的准确性。采用这种技术时,总剂量的大约一半用于一次中心投影,其余剂量则分配到其他断层合成投影视图中。这种采集方法将产生一个投影视图和一个重建视图。本研究的目的之一是评估与均匀剂量(UD)-DBT的三维图像相比,VD采集的单独中心投影是否能提供同等或更好的MC检测效果。另一个目的是比较VD-DBT和UD-DBT的三维重建的肿块检测能力。在一项关于MC检测的定位接收器操作特性(LROC)观察者研究中,作者将VD采集方案(剂量为2 mGy)的中心投影(探测器像素尺寸为100微米)与UD-DBT重建(剂量为4 mGy)(体素尺寸为100微米)进行了比较。研究中使用了尺寸为150和180微米的MC,每个簇由七个随机分布在小体积内的MC组成。UD-DBT中的重建图像是从总剂量为4 mGy的投影集中获得的。当前研究表明,对于MC检测,仅使用VD采集方案的中心投影时,LROC曲线下面积(AL)为0.76,比使用UD采集方案的三维重建图像(AL = 0.84)表现更差。二维方差分析发现在0.05的显著性水平上存在统计学显著差异(p = 0.038)。在当前研究中,虽然也可以使用VD采集方案获得重建图像,但它未用于辅助仅使用中心投影完成的MC检测任务。在评估肿块检测准确性的情况下,将VD-DBT的重建(AL = 0.71)与UD-DBT获得的重建(AL = 0.78)进行了比较。作者发现两者之间没有统计学显著差异(p值 = 0.22),尽管所有观察者对UD-DBT的表现都更好。