Prince Harry E, Lopez Janice, Yeh Cindy, Tablante Joselita, Morgan Joseph, Kaneko Beverly, Duffey Paul
Focus Diagnostics, Inc., Cypress, CA 90630, USA.
Diagn Microbiol Infect Dis. 2009 Oct;65(2):99-102. doi: 10.1016/j.diagmicrobio.2009.06.002.
Brucella IgG and IgM ELISA kits manufactured by Euroimmun (Lubeck, Germany) were evaluated in a reference laboratory setting. Intraassay coefficient of variation (CV) values were < or =10% for positive sera and < or =12% for negative sera; interassay CVs were < or =12% for positive sera and < or =20% for negative sera. The tube agglutination test (TAT) was performed on 51 sera exhibiting various ELISA reactivity profiles. All 18 sera negative for both IgG and IgM by ELISA were TAT negative (titer <1:80), whereas 31 (94%) of 33 sera positive for IgG and/or IgM were TAT positive; the 2 discordant sera were IgG positive IgM negative by ELISA. None of 41 sera from healthy laboratory employees were ELISA IgG positive, whereas 1 (2%) of 41 was ELISA IgM positive. Similarly, 0 of 149 potentially cross-reactive sera (containing rheumatoid factor or antibodies to selected Gram-negative bacteria) was ELISA IgG positive, whereas 4 (3%) of 149 were ELISA IgM positive. These findings demonstrate the acceptable performance of the Euroimmun ELISAs for Brucella antibodies.
对德国吕贝克欧蒙公司生产的布鲁氏菌IgG和IgM酶联免疫吸附测定(ELISA)试剂盒在一个参考实验室环境中进行了评估。阳性血清的批内变异系数(CV)值≤10%,阴性血清的批内CV值≤12%;阳性血清的批间CV值≤12%,阴性血清的批间CV值≤20%。对51份呈现不同ELISA反应性谱的血清进行了试管凝集试验(TAT)。ELISA检测IgG和IgM均为阴性的所有18份血清TAT均为阴性(滴度<1:80),而IgG和/或IgM呈阳性的33份血清中有31份(94%)TAT呈阳性;2份不一致的血清ELISA检测为IgG阳性IgM阴性。41名健康实验室工作人员的血清中无一例ELISA检测IgG呈阳性,而41例中有1例(2%)ELISA检测IgM呈阳性。同样,149份可能存在交叉反应的血清(含类风湿因子或针对选定革兰氏阴性菌的抗体)中无一例ELISA检测IgG呈阳性,而149份中有4份(3%)ELISA检测IgM呈阳性。这些结果表明欧蒙布鲁氏菌抗体ELISA试剂盒性能良好。