Sadeghi Mahdi, Hamed Hosseini S
Agricultural, Medical and Industrial Research School, Nuclear Science and Technology Research Institute, P.O. Box 31485-498, Gohardasht, Karaj, Iran.
Appl Radiat Isot. 2010 Jan;68(1):211-3. doi: 10.1016/j.apradiso.2009.08.007. Epub 2009 Aug 25.
Monte Carlo calculations have been performed using the MCNP4C code for an iodine seed design. As the ADVANTAGE I-125, Model IAI-125 source is commercially available for interstitial brachytherapy treatment, dosimetric characteristics (dose rate constant, radial dose function and anisotropy function) of this source were theoretically determined following the updated AAPM task group 43 (TG-43U1) recommendations. The dose distribution around the seed was calculated with Monte Carlo simulation in liquid water. The Monte Carlo calculated dose rate constant of this source in water was found to be 0.986cGyh(-1)U(-1), with an approximate uncertainty of 0.4%. The obtained result has been compared with the previous study. Comparison of the calculated dose rate constant with the value presented by Meigooni et al. shows a very good agreement. Also the anisotropy function and the radial dose function for this source are graphically compared.
已使用MCNP4C代码对碘籽源设计进行了蒙特卡罗计算。由于ADVANTAGE I - 125型IAI - 125源可用于间质近距离放射治疗,根据最新的美国医学物理学家协会任务组43(TG - 43U1)建议,从理论上确定了该源的剂量学特征(剂量率常数、径向剂量函数和各向异性函数)。在液态水中通过蒙特卡罗模拟计算了籽源周围的剂量分布。该源在水中的蒙特卡罗计算剂量率常数为0.986cGyh(-1)U(-1),近似不确定度为0.4%。将所得结果与先前的研究进行了比较。计算得到的剂量率常数与Meigooni等人给出的值进行比较,显示出非常好的一致性。还对该源的各向异性函数和径向剂量函数进行了图形比较。