Suppr超能文献

电化学沉积纳米羟基磷灰石对喷砂/双酸蚀钛种植体骨结合的影响。

Effect of electrochemically deposited nanohydroxyapatite on bone bonding of sandblasted/dual acid-etched titanium implant.

机构信息

Department of Implantology, Stomatology Hospital, School of Medicine, Zhejiang University, Hangzhou, China.

出版信息

Int J Oral Maxillofac Implants. 2009 Sep-Oct;24(5):790-9.

Abstract

PURPOSE

The aim of this study was to evaluate the effect of an electrochemically deposited nanohydroxyapatite (EDHA) coating on the bone bonding of sandblasted and dual acid-etched titanium implants.

MATERIALS AND METHODS

One hundred EDHA-coated and uncoated sandblasted/dual acid-etched implants (3 mm in diameter, 10 mm long) were inserted into the femoral condyles of 50 rabbits. The osteotomy sites were enlarged to 3 mm in diameter via a sequence of drills. After 2, 4, 6, 8, and 12 weeks of bone healing, removal torque testing was performed to evaluate the interfacial shear strength of each implant type. The removed implants were prepared and observed with an electron microscope equipped with an energy dispersive electron probe x-ray microanalyzer.

RESULTS

The mean removal torque values for the EDHA-coated implants were 39.6 Ncm at 2 weeks and 40.4 Ncm at 4 weeks; corresponding values for the control implants were 21.1 Ncm and 24.1 Ncm. Removal torque values of the EDHA-coated implants were 87% higher than those of control implants after 2 weeks of healing (P = .015). However, the mean removal torque values for both types of implants were similar after 6, 8, or 12 weeks of healing (no significant differences between the implant surfaces; P > .05).

CONCLUSIONS

The EDHA nanocrystal coating had a beneficial effect on interfacial shear strength during the early stages of bone healing.

摘要

目的

本研究旨在评估电化学沉积纳米羟基磷灰石(EDHA)涂层对喷砂酸蚀钛种植体骨结合的影响。

材料和方法

将 100 个 EDHA 涂层和未涂层喷砂酸蚀(3 毫米直径,10 毫米长)植入物插入 50 只兔子的股骨髁。通过一系列钻头将骨切迹扩大至 3 毫米直径。在骨愈合 2、4、6、8 和 12 周后,进行拔出扭矩测试,以评估每种植入物类型的界面剪切强度。取出的植入物用配备能量色散电子探针 X 射线微分析仪的电子显微镜进行制备和观察。

结果

EDHA 涂层植入物的平均拔出扭矩值在 2 周时为 39.6 Ncm,在 4 周时为 40.4 Ncm;对照植入物的相应值分别为 21.1 Ncm 和 24.1 Ncm。在愈合 2 周后,EDHA 涂层植入物的拔出扭矩值比对照植入物高 87%(P =.015)。然而,在愈合 6、8 或 12 周后,两种类型植入物的平均拔出扭矩值相似(植入物表面之间无显著差异;P >.05)。

结论

EDHA 纳米晶体涂层在骨愈合的早期阶段对界面剪切强度有有益的影响。