Removal torque values of titanium implants in the maxilla of miniature pigs.
作者信息
Buser D, Nydegger T, Hirt H P, Cochran D L, Nolte L P
机构信息
Department of Oral Surgery, School of Dental Medicine, University of Berne, Switzerland.
出版信息
Int J Oral Maxillofac Implants. 1998 Sep-Oct;13(5):611-9.
The purpose of this study was to compare side-by-side two different titanium screw-type implants in the maxillae of miniature pigs. The test implants had a machined and acid-etched surface (Osseotite) whereas the control implants were sandblasted and acid-etched (SLA). After 4, 8, and 12 weeks of healing, removal torque testing was performed to evaluate the shear strength of the bone-implant interface for both implant types. The results demonstrated significant differences between both implant types (P < .01). Osseotite implants revealed mean removal torque values (RTV) of 62.5 Ncm at 4 weeks, 87.6 Ncm at 8 weeks, and 95.7 Ncm at 12 weeks of healing. In contrast, the SLA implants demonstrated mean RTV of 109.6 Ncm, 196.7 Ncm, and 186.8 Ncm at corresponding healing periods. The mean RTV for SLA implants was 75% to 125% higher than for Osseotite implants up to 3 months of healing.
本研究的目的是在小型猪的上颌骨中并排比较两种不同的钛螺钉型种植体。测试种植体具有机械加工和酸蚀表面(骨结合钛),而对照种植体为喷砂和酸蚀表面(SLA)。在愈合4周、8周和12周后,进行拆除扭矩测试以评估两种种植体类型的骨-种植体界面的剪切强度。结果显示两种种植体类型之间存在显著差异(P <.01)。骨结合钛种植体在愈合4周时的平均拆除扭矩值(RTV)为62.5 Ncm,8周时为87.6 Ncm,12周时为95.7 Ncm。相比之下,SLA种植体在相应愈合期的平均RTV分别为109.6 Ncm、196.7 Ncm和186.8 Ncm。在长达3个月的愈合期内,SLA种植体的平均RTV比骨结合钛种植体高75%至125%。