Reproducibility of optic nerve head topography measurements in eyes with undilated pupils.

机构信息

Department of Ophthalmology, University of California, San Diego, California, U.S.A.

出版信息

J Glaucoma. 1993 Summer;2(2):104-9.

Abstract

A confocal scanning diode laser ophthalmoscope (Retina Tomograph, Heidelberg Engineering, GmbH, Heidelberg, F.R.G.) was used to examine the optic nerve head in both normal healthy (n = 10) and glaucomatous (n = 10) eyes. An image series was obtained from 32 transverse optical sections taken at consecutive height planes (50-80 mum each) over a 1.5-2.5-mm scan depth. Each image was then analyzed to create a topographic map containing 256 +/- 256 pixels (65,536 pixels) with afield of view of 10+/-. Images were obtained through undilated pupils. Reproducibility was 30.1 +/- 7.0 and 31.8 +/- 10.6 mum in normal and glaucomatous eyes, respectively. The instrument provides highly reproducible topographic data of the optic nerve head in patients with undilated pupils.

摘要

使用共焦扫描二极管激光检眼镜(海德堡工程公司,德国海德堡)检查了正常健康眼(n = 10)和青光眼眼(n = 10)的视神经头。从连续高度平面(每次 50-80 µm)上的 32 个横向光学截面获得图像序列,扫描深度为 1.5-2.5 mm。然后对每个图像进行分析,创建一个包含 256 +/- 256 个像素(65,536 个像素)的地形图,视野为 10+/-。图像是通过未散瞳的瞳孔获得的。正常眼和青光眼眼中的重现性分别为 30.1 +/- 7.0 和 31.8 +/- 10.6 µm。该仪器为未散瞳瞳孔的患者提供了高度可重复的视神经头地形图数据。