Bronzert T J, Brewer H B
Clin Chem. 1977 Nov;23(11):2089-98.
A method is described for the reliable, fast, and relatively inexpensive fractionation of plasma lipoproteins and quantitation of their cholesterol content. This procedure requires 350 microliter of plasma and can be completed within 3 h. Plasma lipoproteins (175 microliter of plasma) were prestained with Fat Red 7B and centrifuged (Beckman Airfuge) at plasma density (d = 1.006 kg/liter) and at a solvent density of 1.060 kg/liter, adjusted by adding solid KBr. Prestained centrifuged samples demonstrated the characteristic elevation of chylomicrons in phenotypes I and V, low-density lipoproteins of phenotype II, very-low-density lipoproteins in phenotype IV and V, and continuum of pink color throughout the centrifuge tube, diagnostic of the floating beta lipoprotein of type III. Centrifuged samples were separated into top and bottom fractions by aspiration. Cholesterol was quantitated with an enzymic oxygen-electrode analyzer (Beckman Cholesterol Analyzer). Correlation coefficients between cholesterol values for plasma from normal hyperlipidemic individuals obtained with the Beckman Analyzer vs. the Technicon AutoAnalyzer II and SMAC systems were 0.977 and 0.973, respectively.
本文描述了一种可靠、快速且相对廉价的血浆脂蛋白分级分离方法及其胆固醇含量的定量方法。该程序需要350微升血浆,可在3小时内完成。血浆脂蛋白(175微升血浆)用脂肪红7B预染色,并在血浆密度(d = 1.006千克/升)和通过添加固体溴化钾调节至1.060千克/升的溶剂密度下离心(贝克曼空气离心机)。预染色的离心样品显示出I型和V型表型中乳糜微粒的特征性升高、II型表型的低密度脂蛋白、IV型和V型表型的极低密度脂蛋白,以及整个离心管中的粉红色连续区域,这是III型漂浮β脂蛋白的诊断特征。通过抽吸将离心样品分离为顶部和底部馏分。用酶促氧电极分析仪(贝克曼胆固醇分析仪)对胆固醇进行定量。使用贝克曼分析仪与Technicon AutoAnalyzer II和SMAC系统获得的正常高脂血症个体血浆胆固醇值之间的相关系数分别为0.977和0.973。