Retina. 2010 Jan;30(1):63-70. doi: 10.1097/IAE.0b013e3181bd2c5a.
The purpose of this study was to quantify photoreceptor outer segment (PROS) length in 27 consecutive patients (30 eyes) with diabetic macular edema using spectral domain optical coherence tomography and to describe the correlation between PROS length and visual acuity.
Three spectral domain-optical coherence tomography scans were performed on all eyes during each session using Cirrus HD-OCT. A prototype algorithm was developed for quantitative assessment of PROS length. Retinal thicknesses and PROS lengths were calculated for 3 parameters: macular grid (6 x 6 mm), central subfield (1 mm), and center foveal point (0.33 mm). Intrasession repeatability was assessed using coefficient of variation and intraclass correlation coefficient. The association between retinal thickness and PROS length with visual acuity was assessed using linear regression and Pearson correlation analyses. The main outcome measures include intrasession repeatability of macular parameters and correlation of these parameters with visual acuity.
Mean retinal thickness and PROS length were 298 mum to 381 microm and 30 microm to 32 mum, respectively, for macular parameters assessed in this study. Coefficient of variation values were 0.75% to 4.13% for retinal thickness and 1.97% to 14.01% for PROS length. Intraclass correlation coefficient values were 0.96 to 0.99 and 0.73 to 0.98 for retinal thickness and PROS length, respectively. Slopes from linear regression analyses assessing the association of retinal thickness and visual acuity were not significantly different from 0 (P > 0.20), whereas the slopes of PROS length and visual acuity were significantly different from 0 (P < 0.0005). Correlation coefficients for macular thickness and visual acuity ranged from 0.13 to 0.22, whereas coefficients for PROS length and visual acuity ranged from -0.61 to -0.81.
Photoreceptor outer segment length can be quantitatively assessed using Cirrus HD-OCT. Although the intrasession repeatability of PROS measurements was less than that of macular thickness measurements, the stronger correlation of PROS length with visual acuity suggests that the PROS measures may be more directly related to visual function. Photoreceptor outer segment length may be a useful physiologic outcome measure, both clinically and as a direct assessment of treatment effects.
本研究的目的是使用谱域光相干断层扫描定量测量 27 例(30 只眼)糖尿病性黄斑水肿患者的光感受器外节(PROS)长度,并描述 PROS 长度与视力之间的相关性。
在每次检查中,所有眼均使用 Cirrus HD-OCT 进行 3 次谱域光相干断层扫描。开发了一种用于定量评估 PROS 长度的原型算法。为 3 个参数计算视网膜厚度和 PROS 长度:黄斑网格(6 x 6mm)、中央子区(1mm)和中心黄斑点(0.33mm)。使用变异系数和组内相关系数评估了日内重复性。使用线性回归和 Pearson 相关分析评估视网膜厚度和 PROS 长度与视力之间的相关性。主要观察指标包括:评估日内黄斑参数的重复性以及这些参数与视力的相关性。
本研究中评估的黄斑参数的平均视网膜厚度和 PROS 长度分别为 298μm 至 381μm 和 30μm 至 32μm。视网膜厚度的变异系数值为 0.75%至 4.13%,PROS 长度的变异系数值为 1.97%至 14.01%。视网膜厚度和 PROS 长度的组内相关系数值分别为 0.96 至 0.99 和 0.73 至 0.98。评估视网膜厚度和视力相关性的线性回归分析的斜率与 0 无显著差异(P>0.20),而 PROS 长度和视力相关性的斜率与 0 有显著差异(P<0.0005)。黄斑厚度和视力的相关系数范围为 0.13 至 0.22,而 PROS 长度和视力的相关系数范围为-0.61 至-0.81。
使用 Cirrus HD-OCT 可以定量评估光感受器外节长度。虽然 PROS 测量的日内重复性小于黄斑厚度测量的重复性,但 PROS 长度与视力的相关性更强,提示 PROS 测量可能与视觉功能更直接相关。光感受器外节长度可能是一种有用的生理结果测量指标,具有临床和直接评估治疗效果的双重作用。