使用心尖窗通过脉冲多普勒测量新生儿的心输出量。

Cardiac output by pulsed Doppler in neonates using the apical window.

作者信息

Mandelbaum-Isken V H, Linderkamp O

机构信息

Division of Neonatology, University of Heidelberg, FRG.

出版信息

Pediatr Cardiol. 1991 Jan;12(1):13-6. doi: 10.1007/BF02238491.

Abstract

In 18 healthy neonates, cardiac output was measured by range-gated Doppler echocardiography with simultaneous two-dimensional echocardiography, using the apical approach. Maximal and modal flow velocities were compared by using spectral analysis of the Doppler signal. Blood flow velocities were measured in the left ventricular outflow tract and in the ascending aorta. Cardiac index (cardiac output per kg body weight), with the sample volume placed in the left ventricular outflow, averaged 170 +/- 45 ml/min/kg in the maximal flow velocity mode and 133 +/- 39 ml/min/kg in the modal mode, respectively (p less than 0.05). Sampling in the ascending aorta gave cardiac index values of 198 +/- 73 ml/min/kg in the maximal velocity and 167 +/- 58 ml/min/kg in the modal velocity mode. Maximal flow velocities in the ascending aorta came closest to published data obtained by invasive and noninvasive methods.

摘要

在18名健康新生儿中,采用心尖途径,通过距离选通多普勒超声心动图结合同步二维超声心动图测量心输出量。使用多普勒信号的频谱分析比较最大流速和众数流速。在左心室流出道和升主动脉中测量血流速度。在心输出量(每千克体重的心输出量)测量中,将取样容积置于左心室流出道,最大流速模式下平均为170±45毫升/分钟/千克,众数模式下为133±39毫升/分钟/千克(p<0.05)。在升主动脉取样时,最大流速下心输出量指数值为198±73毫升/分钟/千克,众数流速模式下为167±58毫升/分钟/千克。升主动脉中的最大流速最接近通过有创和无创方法获得的已发表数据。