Pediatric radiography entrance doses for some routine procedures in three hospitals within eastern Nigeria.

作者信息

Egbe N O, Inyang S O, Ibeagwa O B, Chiaghanam N O

机构信息

Department of Radiography, University of Calabar, Nigeria.

出版信息

J Med Phys. 2008 Jan;33(1):29-34. doi: 10.4103/0971-6203.39422.

Abstract

A survey of the entrance surface doses in the routine radiography of children in eastern Nigeria has been carried out in three hospitals, using thermoluminescence detectors. Chest, abdomen, lumbar spine, skull and pelvis were covered in this study. Findings reveal that doses are higher than the recommended reference values elsewhere, as well as values reported for Sudan. The mean percentage difference in entrance doses for chest radiography for this study and an earlier one carried out for three hospitals in the west of Nigeria is about 44.7%. The high doses are traceable to a lack of standardization in procedure, resulting in use of low tube voltages and high currents for examination, as well as the status of facilities in the area. Recommendations are made for immediate corrective measures to lower the doses.

摘要

在尼日利亚东部的三家医院,使用热释光探测器对儿童常规放射摄影中的体表入射剂量进行了一项调查。本研究涵盖了胸部、腹部、腰椎、颅骨和骨盆。研究结果显示,这些剂量高于其他地方推荐的参考值,也高于苏丹报告的值。本研究与早前在尼日利亚西部三家医院进行的一项研究相比,胸部放射摄影体表入射剂量的平均百分比差异约为44.7%。高剂量可归因于程序缺乏标准化,导致检查时使用低管电压和高电流,以及该地区的设备状况。建议立即采取纠正措施以降低剂量。