使用 TLD 评估约旦北部三家公立医院常规诊断 X 射线检查中的患者辐射剂量。

Assessment of Patient Radiation Dose during Conventional Diagnostic X-ray Examinations in Three Public Hospitals in Northern Jordan using TLDs.

机构信息

Physics Department, Faculty of Science, Yarmouk University, Irbid, Jordan.

出版信息

Health Phys. 2020 Jun;118(6):593-599. doi: 10.1097/HP.0000000000001155.

Abstract

This study aims to measure entrance surface doses during routine chest and abdomen x-ray examinations of adult and child patients. Radiation dose measurements were performed using thermoluminescent dosimeters TLD-100s in three major public hospitals in northern Jordan on a total of 100 patients. Wide variations in entrance surface doses were observed within and between hospitals, which might be attributed to significant variations of the selected exposure parameters. For adult patients, the results have shown that the majority of entrance surface dose values from both chest and abdomen examinations were within recommended values of diagnostic reference levels. For child patients, the mean entrance surface dose from chest examinations in three age groups were 0.131 mGy (0-1 y), 0.136 mGy (1-5 y), and 0.191 mGy (5-10 y). These values were considered relatively high compared to the European reference levels and published results in the literature. However, for abdomen examinations, entrance surface dose values were relatively lower than European reference levels. Patient effective doses were estimated using a PCXMC 2.0 Monte Carlo program. The results for both adults and children were found to be relatively lower than the values reported by international publications. Due to the wide variations of entrance surface dose and the higher radiation doses delivered to child patients, this study recommends implementing a quality assurance program in such hospitals to achieve optimization between good image quality and minimum dose according to the as low as reasonably achievable principle. Moreover, the results of this work will provide a useful base for establishing local diagnostic reference levels for chest and abdomen examinations in Jordan.

摘要

本研究旨在测量成人和儿童患者常规胸部和腹部 X 射线检查的体表入射剂量。在约旦北部的三家主要公立医院,使用 TLD-100 型热释光剂量计对总共 100 名患者进行了辐射剂量测量。在医院内和医院之间观察到体表入射剂量存在广泛的差异,这可能归因于所选曝光参数的显著差异。对于成人患者,结果表明,来自胸部和腹部检查的大多数体表入射剂量值均在诊断参考水平的推荐值范围内。对于儿童患者,三组年龄的胸部检查的平均体表入射剂量分别为 0.131 mGy(0-1 岁)、0.136 mGy(1-5 岁)和 0.191 mGy(5-10 岁)。与欧洲参考水平和文献中公布的结果相比,这些值被认为相对较高。然而,对于腹部检查,体表入射剂量值相对低于欧洲参考水平。使用 PCXMC 2.0 蒙特卡罗程序估算了患者有效剂量。结果表明,成人和儿童的剂量均低于国际出版物报告的值。由于体表入射剂量的广泛差异以及儿童患者接受的辐射剂量较高,因此本研究建议在这些医院实施质量保证计划,以根据尽可能低的原则实现良好的图像质量和最小剂量之间的优化。此外,这项工作的结果将为在约旦建立胸部和腹部检查的本地诊断参考水平提供有用的基础。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验