多叶准直器的射频噪声:在直线加速器磁共振系统中使用多叶准直器的可行性研究。

Radio frequency noise from an MLC: a feasibility study of the use of an MLC for linac-MR systems.

机构信息

Department of Physics, University of Alberta, Alberta, T6G 2G7, Canada.

出版信息

Phys Med Biol. 2010 Feb 21;55(4):981-94. doi: 10.1088/0031-9155/55/4/005. Epub 2010 Jan 20.

Abstract

Currently several groups are actively researching the integration of a megavoltage teletherapy unit with magnetic resonance (MR) imaging for real-time image-guided radiotherapy. The use of a multileaf collimator (MLC) for intensity-modulated radiotherapy for linac-MR units must be investigated. The MLC itself will likely reside in the fringe field of the MR and the motors will produce radio frequency (RF) noise. The RF noise power spectral density from a Varian 52-leaf MLC motor, a Varian Millennium MLC motor and a brushless fan motor has been measured as a function of the applied magnetic field using a near field probe set. For the Varian 52-leaf MLC system, the RF noise produced by 13 of 52 motors is studied as a function of distance from the MLC. Data are reported in the frequency range suitable for 0.2-1.5 T linac-MR systems. Below 40 MHz the Millennium MLC motor tested showed more noise than the Varian 52-leaf motor or the brushless fan motor. The brushless motor showed a small dependence on the applied magnetic field. Images of a phantom were taken by the prototype linac-MR system with the MLC placed in close proximity to the magnet. Several orientations of the MLC in both shielded and non-shielded configurations were studied. For the case of a non-shielded MLC and associated cables, the signal-to-noise ratio (SNR) was reduced when 13 of 52 MLC leaves were moved during imaging. When the MLC and associated cables were shielded, the measured SNR of the images with 13 MLC leaves moving was experimentally the same as the SNR of the stationary MLC image. When the MLC and cables are shielded, subtraction images acquired with and without MLC motion contains no systematic signal. This study illustrates that the small RF noise produced by functioning MLC motors can be effectively shielded to avoid SNR degradation. A functioning MLC can be incorporated into a linac-MR unit.

摘要

目前,有几个研究小组正在积极研究将兆伏级放射治疗设备与磁共振(MR)成像相结合,实现实时图像引导放射治疗。需要对直线加速器-MR 单元的多叶准直器(MLC)进行强度调制放射治疗的应用研究。MLC 本身可能位于 MR 的边缘场中,电机将产生射频(RF)噪声。使用近场探头组测量了瓦里安 52 叶 MLC 电机、瓦里安千禧 MLC 电机和无刷风扇电机的 RF 噪声功率谱密度随外加磁场的变化。对于瓦里安 52 叶 MLC 系统,研究了 52 个电机中的 13 个电机产生的 RF 噪声随距 MLC 的距离的变化。在适用于 0.2-1.5T 直线加速器-MR 系统的频率范围内报告数据。在 40MHz 以下,测试的千禧 MLC 电机的噪声大于瓦里安 52 叶电机或无刷风扇电机。无刷电机对外加磁场的依赖性较小。通过原型直线加速器-MR 系统拍摄了一个模体的图像,将 MLC 放置在靠近磁铁的位置。研究了在屏蔽和非屏蔽配置下 MLC 的几种取向。对于非屏蔽的 MLC 和相关电缆的情况,当在成像过程中移动 52 个 MLC 叶片中的 13 个时,信号噪声比(SNR)降低。当 MLC 和相关电缆被屏蔽时,测量到的移动 13 个 MLC 叶片的图像 SNR 与静止 MLC 图像的 SNR 相同。当 MLC 和电缆被屏蔽时,在有无 MLC 运动的情况下采集的差影图像没有系统信号。这项研究表明,运行中的 MLC 电机产生的小 RF 噪声可以有效地屏蔽,以避免 SNR 下降。一个运行中的 MLC 可以被整合到直线加速器-MR 单元中。

相似文献

1
Radio frequency noise from an MLC: a feasibility study of the use of an MLC for linac-MR systems.
Phys Med Biol. 2010 Feb 21;55(4):981-94. doi: 10.1088/0031-9155/55/4/005. Epub 2010 Jan 20.
2
Brushed permanent magnet DC MLC motor operation in an external magnetic field.
Med Phys. 2010 May;37(5):2131-4. doi: 10.1118/1.3392165.
3
Performance of a multileaf collimator system for a 1.5T MR-linac.
Med Phys. 2021 Feb;48(2):546-555. doi: 10.1002/mp.14608. Epub 2020 Dec 14.
4
Impact of the MLC on the MRI field distortion of a prototype MRI-linac.
Med Phys. 2013 Dec;40(12):121705. doi: 10.1118/1.4828792.
5
Radio frequency shielding for a linac-MRI system.
Phys Med Biol. 2010 Feb 21;55(4):995-1006. doi: 10.1088/0031-9155/55/4/006. Epub 2010 Jan 20.
6
MRI quality control for low-field MR-IGRT systems: Lessons learned.
J Appl Clin Med Phys. 2019 Oct;20(10):53-66. doi: 10.1002/acm2.12713. Epub 2019 Sep 21.
7
Performance of a multi leaf collimator system for MR-guided radiation therapy.
Med Phys. 2017 Dec;44(12):6504-6514. doi: 10.1002/mp.12571. Epub 2017 Oct 19.
9

引用本文的文献

1
Glamour of technology.
J Med Phys. 2010 Jul;35(3):129-30. doi: 10.4103/0971-6203.66759.

本文引用的文献

3
Radio frequency noise from clinical linear accelerators.
Phys Med Biol. 2009 Apr 21;54(8):2483-92. doi: 10.1088/0031-9155/54/8/015. Epub 2009 Apr 1.
4
A new way of adapting IMRT delivery fraction-by-fraction to cater for variable intrafraction motion.
Phys Med Biol. 2008 Sep 21;53(18):5177-91. doi: 10.1088/0031-9155/53/18/022. Epub 2008 Aug 26.
5
Management of three-dimensional intrafraction motion through real-time DMLC tracking.
Med Phys. 2008 May;35(5):2050-61. doi: 10.1118/1.2905355.
6
Patient dosimetry for hybrid MRI-radiotherapy systems.
Med Phys. 2008 Mar;35(3):1019-27. doi: 10.1118/1.2839104.
7
A (short) history of image-guided radiotherapy.
Radiother Oncol. 2008 Jan;86(1):4-13. doi: 10.1016/j.radonc.2007.11.023.
8
MRI/linac integration.
Radiother Oncol. 2008 Jan;86(1):25-9. doi: 10.1016/j.radonc.2007.10.034. Epub 2007 Nov 26.
9
Motion effects in (intensity modulated) radiation therapy: a review.
Phys Med Biol. 2006 Jul 7;51(13):R403-25. doi: 10.1088/0031-9155/51/13/R23. Epub 2006 Jun 20.
10
IMRT: a review and preview.
Phys Med Biol. 2006 Jul 7;51(13):R363-79. doi: 10.1088/0031-9155/51/13/R21. Epub 2006 Jun 20.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验