Department of Orthodontics, University Medical Center Groningen, University of Groningen, PO Box 30.001, Groningen, 9700 RB, The Netherlands.
Am J Orthod Dentofacial Orthop. 2010 Jan;137(1):16.e1-6; discussion 16-7. doi: 10.1016/j.ajodo.2009.06.016.
The aims of this study were to determine the linear accuracy of 3-dimensional surface models derived from a commercially available cone-beam computed tomography (CBCT) dental imaging system and volumetric rendering software and to investigate the influence of voxel resolution on the linear accuracy of CBCT surface models.
Glass sphere markers were fixed on 10 dry mandibles. The mandibles were scanned with 0.40 and 0.25 voxel size resolutions in 3 sessions. Anatomic truth was established with 6 direct digital caliper measurements. The surface models were rendered by a volumetric rendering program, and the CBCT measurements were established as the mean of the 3 measurements.
The intraclass correlation coefficients between the physical measurements and the measurements of the CBCT images of 0.40 and 0.25 voxels were all more than 0.99. All CBCT measurements were accurate. There was no difference between the accuracy of the measurements between the 0.40 and 0.25 voxel size groups. The smallest detectable differences of the CBCT measurements were minimal, confirming the accuracy of the CBCT measurement procedure.
The measurements on 3-dimensional surface models of 0.25 and 0.40 voxel size data sets made with the 3D eXam CBCT scanner (KaVo Dental GmbH, Bismarckring, Germany) and SimPlant Ortho Pro software (version 2.00, Materialise Dental, Leuven, Belgium) are accurate compared with direct caliper measurements. An increased voxel resolution did not result in greater accuracy of the surface model measurments.
本研究旨在确定一种商用锥形束 CT(CBCT)牙科成像系统和体绘制软件得出的三维表面模型的线性精度,并研究体素分辨率对 CBCT 表面模型线性精度的影响。
将玻璃球标记物固定在 10 个干燥的下颌骨上。下颌骨以 0.40 和 0.25 体素分辨率分 3 次扫描。使用 6 个直接数字卡尺测量确定解剖学真值。通过体绘制程序渲染表面模型,将 CBCT 测量值定义为 3 次测量的平均值。
0.40 和 0.25 体素的物理测量值与 CBCT 图像测量值之间的组内相关系数均大于 0.99。所有 CBCT 测量值均准确。0.40 和 0.25 体素组之间的测量精度无差异。CBCT 测量的最小可检测差异最小,证实了 CBCT 测量程序的准确性。
与直接卡尺测量相比,使用 3D eXam CBCT 扫描仪(德国卡瓦齿科有限公司,俾斯麦环)和 SimPlant Ortho Pro 软件(版本 2.00,比利时 Materialise Dental)得出的 0.25 和 0.40 体素数据集的三维表面模型测量值准确。体素分辨率的增加不会导致表面模型测量值更准确。