Suppr超能文献

不同体素大小和不同分割阈值设置的 CBCT 在下颌容积测量中的体外准确性。

Accuracy of in vitro mandibular volumetric measurements from CBCT of different voxel sizes with different segmentation threshold settings.

机构信息

Department of Orthodontics, Ninth People's Hospital Affiliated to Shanghai Jiao Tong University, School of Medicine, No. 639 Zhizaoju Road, Shanghai, China.

出版信息

BMC Oral Health. 2019 Sep 4;19(1):206. doi: 10.1186/s12903-019-0891-5.

Abstract

BACKGROUND

To determine the accuracy of volumetric measurements of the mandible in vitro by cone-beam computed tomography (CBCT) and to analyze the influence of voxel sizes and segmentation threshold settings on it.

METHODS

The samples were obtained from pig mandibles and scanned with 4 voxel sizes: .125 mm, .20 mm, .30 mm, and .40 mm. The minimum segmentation thresholds in Hounsfield units (HU) were set as 0, 100, 200, 300, and 400, respectively, for each voxel size for 3D reconstruction. Laser scanning as the reference, the volumes of each CBCT scanning, the mean iterative distances of superimposition and total positive and negative deviations were recorded and compared.

RESULTS

The volumes of CBCT-scan deviated from those of laser-scan by + 7.67% to - 3.05% with different HU and voxel sizes. The deviation increased with the voxel size. There was a more suitable minimum HU threshold of segmentation (HU100 for .125 mm, 200 for .20 mm, 300 for .30 mm, and 400 for .40 mm) for each voxel size.

CONCLUSIONS

Voxel sizes and Hounsfield unit thresholds influence the accuracy of volumetric measurements in CBCT scanning. The volume increase with the voxel size, and different voxel sizes correspond to different optimal Hounsfield unit thresholds.

摘要

背景

通过锥形束 CT(CBCT)确定下颌骨的体外容积测量的准确性,并分析体素大小和分割阈值设置对其的影响。

方法

样本取自猪下颌骨,并以 4 个体素大小进行扫描:0.125mm、0.20mm、0.30mm 和 0.40mm。对于每个体素大小的 3D 重建,分别将亨氏单位(HU)的最小分割阈值设置为 0、100、200、300 和 400。记录并比较了激光扫描作为参考的每个 CBCT 扫描的体积、叠加的平均迭代距离和总正负偏差。

结果

CBCT 扫描的体积与激光扫描的体积相差+7.67%至-3.05%,不同的 HU 和体素大小都有差异。偏差随体素大小而增加。对于每个体素大小,存在更合适的分割最小 HU 阈值(HU100 用于 0.125mm,200 用于 0.20mm,300 用于 0.30mm,400 用于 0.40mm)。

结论

体素大小和亨氏单位阈值会影响 CBCT 扫描中容积测量的准确性。体积随体素大小而增加,不同的体素大小对应不同的最佳亨氏单位阈值。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/aaa4/6727515/4dfadf956b15/12903_2019_891_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验