STUDY OBJECTIVE

To determine the difference in performance of two different auditory evoked potentials (AEP) monitors, the A-Line AEP (AAI) and the aepEX, and their indices, during general anesthesia.

DESIGN

Prospective study.

SETTING

Operating room at a private hospital.

PATIENTS

40 ASA physical status I and II women, aged 30 to 70 years, scheduled for partial mastectomy.

INTERVENTIONS

Anesthesia was induced with propofol and fentanyl, and a Laryngeal Mask Airway (LMA) was inserted. Anesthesia was maintained with propofol, fentanyl, and nitrous oxide.

MEASUREMENTS AND MAIN RESULTS

The AAI or the aepEX was continuously monitored and their performance was compared at the start of monitoring, at LMA insertion, after disturbance by electric cautery, and during anesthesia. Eighteen of 20 patients had low enough impedance to extract good electroencephalogram signals at the first electrode application with the A-Line AEP, and 14 of 20 patients, with the aepEX. The time to return to good signals after signal disturbance by electric cautery was 14 +/- 3 seconds (SD) with the AAI and 19 +/- 4 seconds (SD) with the aepEX (P = 0.035). Both AAI and aepEX decreased after anesthesia induction, with significantly lower values seen in AAI than the aepEX.

CONCLUSIONS

The A-Line AEP (AAI) is better detects the response to painful stimuli and during recovering from noise of electric cautery than the aepEX. The aepEX shows higher values than the AAI during propofol-fentanyl-nitrous oxide anesthesia.

摘要

研究目的

确定两种不同的听觉诱发电位(AEP)监测仪,即 A-Line AEP(AAI)和 aepEX,及其指数在全身麻醉期间的性能差异。

设计

前瞻性研究。

地点

私立医院的手术室。

患者

40 名 ASA 身体状况 I 和 II 的女性,年龄 30 至 70 岁,计划行乳房部分切除术。

干预措施

麻醉诱导使用丙泊酚和芬太尼,并插入喉罩气道(LMA)。使用丙泊酚、芬太尼和氧化亚氮维持麻醉。

测量和主要结果

连续监测 AAI 或 aepEX,并在监测开始时、LMA 插入时、电灼干扰后以及麻醉期间比较其性能。20 名患者中有 18 名患者的阻抗足够低,第一次使用 A-Line AEP 时可以提取出良好的脑电图信号,而 20 名患者中有 14 名患者可以使用 aepEX。电灼干扰后信号恢复良好的时间为 AAI 为 14±3 秒(SD),aepEX 为 19±4 秒(SD)(P=0.035)。AEP 诱导后 AAI 和 aepEX 均降低,AAI 值明显低于 aepEX。

结论

与 aepEX 相比,A-Line AEP(AAI)更能检测到对疼痛刺激的反应,并在从电灼噪声中恢复时更好。在丙泊酚-芬太尼-氧化亚氮麻醉期间,aepEX 的值高于 AAI。