Liedtke R J, Busby B, Batjer J D
Clin Chem. 1978 Jan;24(1):161-5.
The enzymatic cholesterol method used with the Du Pont aca has been modified to provide a reliable measurement of high-density lipoprotein cholesterol in serum after heparin/Mn2+ precipitation of the low- and very-low-density lipoproteins. Interference by Mn2+, equivalent to about 90 mg of cholesterol per liter, is decreased to less than 40 mg of cholesterol per liter by the presence of ethylenediaminetetraacetate (8 mmol/liter) in the diluent; the residual effect of Mn2+ is compensated by calibrating the aca with standards containing Mn2+ and heparin. With an 80-microliter sample, the sensitivity is 236 muA/mg per liter and linearity ranges from 50 to 1500 mg/liter. Average analytical recovery of cholesterol added to the high-density lipoprotein fraction was 103%. Diluted fractions give the expected results. Between-run reproducibility (CV) is 1.3 and 1.6% at 463 and 554 mg/liter. Correlation with the Lipid Research Clinics' procedure (25 samples) gave a regression line of y(aca) = 1.039x- 15, and a correlation coefficient of 0.997.
与杜邦aca一起使用的酶法胆固醇测定方法已作改进,以便在通过肝素/Mn2+沉淀法去除血清中的低密度和极低密度脂蛋白后,可靠地测定血清中的高密度脂蛋白胆固醇。通过在稀释液中加入乙二胺四乙酸(8毫摩尔/升),可将相当于每升约90毫克胆固醇的Mn2+干扰降低至每升低于40毫克胆固醇;通过用含Mn2+和肝素的标准品校准aca来补偿Mn2+的残留影响。对于80微升的样品,灵敏度为每升236微安/毫克,线性范围为50至1500毫克/升。添加到高密度脂蛋白组分中的胆固醇的平均分析回收率为103%。稀释后的组分给出预期结果。在463和554毫克/升时,批间重复性(CV)分别为1.3%和1.6%。与脂质研究临床中心方法(25个样本)的相关性得出回归直线y(aca)=1.039x - 15,相关系数为0.997。