Department of Pathology, University of Mississippi Medical Center, Jackson, MS 39232, USA.
Clin Chem Lab Med. 2010 Apr;48(4):519-21. doi: 10.1515/CCLM.2010.092.
Random-access analyzers that employ dedicated probes may be affected by carryover. Sample carryover, involving analyte from one sample that is measured in a subsequent sample, is most often detected in measurement procedures that have wide reportable ranges. However, reagent carryover can be more difficult to detect as it may involve only one pair of measurement procedures.
Our laboratory noted several patient samples with total cholesterol <2.58 mmol/L (100 mg/dL) during the initial months after a new chemistry analyzer was installed. The problem seemed to occur intermittently, approximately once per week.
Immediate reanalysis of affected samples resulted in measured values that were 1.03-2.58 mmol/L (40-100 mg/dL) higher. Evaluation of reagent carryover revealed a significant decrease in total cholesterol after analysis of creatine kinase (EC 2.7.3.2). Carryover disappeared when an additional reagent probe wash was applied, whereas the root of the problem was eliminated with replacement of the reagent probes. The insidious nature of reagent carryover made it difficult to initially detect the problem, which affected two of the three instruments in our laboratory.
Laboratorians should be aware of the potential for carryover in random access analyzers, and should be mindful of appropriate troubleshooting techniques.
采用专用探头的随机存取分析仪可能会受到携带污染的影响。样品携带污染,即一个样品中的分析物在下一个样品中被测量,最常发生在报告范围较宽的测量程序中。然而,试剂携带污染可能更难检测,因为它可能只涉及一对测量程序。
我们实验室在新的化学分析仪安装后的最初几个月里,发现了几个总胆固醇<2.58mmol/L(100mg/dL)的患者样本。这个问题似乎间歇性地出现,大约每周一次。
受影响样本的即时重新分析结果显示,测量值高出 1.03-2.58mmol/L(40-100mg/dL)。对试剂携带污染的评估显示,在分析肌酸激酶(EC 2.7.3.2)后,总胆固醇显著下降。当应用额外的试剂探针清洗时,携带污染消失,而更换试剂探针则消除了问题的根源。试剂携带污染的隐匿性质使得最初很难检测到这个问题,它影响了我们实验室的三台仪器中的两台。
实验室人员应该意识到随机存取分析仪中存在携带污染的可能性,并注意适当的故障排除技术。