Centre for Medical Radiation Physics, University of Wollongong, Gwynneville, NSW, Australia.
Phys Med. 2010 Oct;26(4):216-9. doi: 10.1016/j.ejmp.2010.01.003. Epub 2010 Feb 10.
Gafchromic XRQA radiochromic film, is an effective tool for quality assurance and dose assessment in kilovoltage radiotherapy and diagnostic applications. Like other Gafchromic film products, XRQA film exhibits a variation in dose to reflected optical density response with angle of rotation when analysed with a light source that is partially or fully polarised such as a desktop scanner. Although warnings are not given on manufacturers specifications, this can affect dosimetry accuracy and we recommend that it is essential to scan all XRQA films in the same orientation. The effect is not as pronounced as EBT Gafchromic film. The magnitude of this variation has been measured and shown to be up to 16 ± 2% (1SD) using a fully linear polarised light source was seen with a 90° angle rotation. This would be the maximum variation seen on a desktop scanner with a fully polarised light source. For our standard desktop scanner (Epson v700) a mean variation of 2 ± 1% from 0 cGy to 20 cGy applied dose was measured as compared to 8 ± 2% for EBT Gafchromic. We recommend that to decrease uncertainty in dose measurement, accurate alignment of the calibration films to experimental films be performed on a regular basis. This is especially important if your desktop scanner has a high degree of polarization of its light source.
Gafchromic XRQA 光致变色胶片是千伏放射治疗和诊断应用中质量保证和剂量评估的有效工具。与其他 Gafchromic 胶片产品一样,XRQA 胶片在使用部分或完全偏振光源(如台式扫描仪)分析时,其剂量与反射光密度响应的旋转角度之间存在变化。尽管制造商规格中没有警告,但这会影响剂量准确性,因此我们建议务必以相同的方向扫描所有 XRQA 胶片。这种影响不如 EBT Gafchromic 胶片明显。已经测量了这种变化的幅度,当使用完全线性偏振光源时,在 90°角度旋转时,最大变化幅度为 16±2%(1SD)。这将是带有完全偏振光源的台式扫描仪上看到的最大变化。对于我们的标准台式扫描仪(Epson v700),与 EBT Gafchromic 的 8±2%相比,在 0 cGy 至 20 cGy 应用剂量下,测量到的平均变化为 2±1%。我们建议为了降低剂量测量的不确定性,应定期将校准胶片与实验胶片进行精确对准。如果您的台式扫描仪的光源具有很高的偏振度,这一点尤为重要。