Dipartimento di Fisiopatologia Clinica, Universita' di Firenze, Firenze, Italy.
J Appl Clin Med Phys. 2010 Jan 29;11(1):3128. doi: 10.1120/jacmp.v11i1.3128.
In radiotherapy treatment portal images and digitally reconstructed radiographs (DRRs) are the most frequently used image-guidance systems to monitor patient set-up during clinical routine. DRR is therefore a critical element in the process of virtual simulation and patient position verification in radiotherapy. The aim of this work was to assess the geometric reliability of the DRR generated by XiO, Pinnacle, Odyssey and Ocentra-Masterplan Treatment Planning Systems (TPS). Instead of the Computed Tomography (CT) images of a quality assurance phantom two synthetic phantoms were created and used to asses the linearity and the ray tracing reliability of the DRR generators. The obtained results demonstrate that the linearity and the reliability of DRR generators was less than 1 pixel for all TPSs.
在放射治疗中,治疗时的射野影像和数字重建射线影像(DRR)是最常使用的影像引导系统,用于在临床常规中监测患者摆位。因此,DRR 是放射治疗中虚拟模拟和患者位置验证过程中的关键要素。这项工作的目的是评估 XiO、Pinnacle、Odyssey 和 Ocentra-Masterplan 治疗计划系统(TPS)生成的 DRR 的几何可靠性。本研究不是使用质量保证体模的 CT 图像,而是创建了两个合成体模来评估 DRR 生成器的线性度和射线追踪可靠性。所得结果表明,所有 TPS 的 DRR 生成器的线性度和可靠性都小于 1 像素。