Leimgruber A, Mosimann B, Claeys M, Seppey M, Jaccard Y, Aubert V, Peitrequin R, Nisoli M P, Pécoud A
Division of Immunology and Allergy, Centre Hospitalier Universitaire Vaudois, Lausanne, Switzerland.
Clin Exp Allergy. 1991 Jan;21(1):127-31. doi: 10.1111/j.1365-2222.1991.tb00814.x.
The Pharmacia CAP system is a new assay for specific IgE characterized by a new solid phase (hydrophilic polymer encased in a capsule). The CAP results were compared to those of the Phadebas IgE RAST and skin-prick tests (SPT) performed in 145 subjects suffering from rhinitis and/or asthma with eight common inhalant allergens (total number of tests = 1160). Concording CAP/RAST results were found in 91% of the tests. The CAP was positive in 78% and the RAST in 65% of the positive SPT. Conversely, the CAP was negative in 90.6% and the RAST in 96.4% of the negative SPT. A pattern negative SPT, negative RAST and positive CAP' was found in 56 tests (40 subjects): in four such subjects, the CAP positivity was confirmed by a positive IgE crossed radioimmunoelectrophoresis. Three borderline positive results were found among 240 negative controls (serum from cord blood or non-atopics). These data indicate that compared with SPT the CAP system is (a) more sensitive than the Pharmacia RAST and (b) does not seem less specific.
法玛西亚CAP系统是一种针对特异性IgE的新检测方法,其特点是采用了一种新的固相(包裹在胶囊中的亲水性聚合物)。对145名患有鼻炎和/或哮喘的受试者进行了针对8种常见吸入性变应原的检测(检测总数 = 1160),并将CAP检测结果与Phadebas IgE RAST检测结果及皮肤点刺试验(SPT)结果进行了比较。在91%的检测中发现CAP/RAST结果一致。在SPT阳性的结果中,CAP阳性率为78%,RAST阳性率为65%。相反,在SPT阴性的结果中,CAP阴性率为90.6%,RAST阴性率为96.4%。在56次检测(40名受试者)中发现了“SPT阴性、RAST阴性但CAP阳性”的模式:在4名此类受试者中,通过IgE交叉放射免疫电泳阳性证实了CAP阳性。在240名阴性对照(脐血或非特应性个体的血清)中发现了3个临界阳性结果。这些数据表明,与SPT相比,CAP系统(a)比法玛西亚RAST更敏感,(b)特异性似乎并不更低。