Department of Surgery, University Hospital Technical University of Aachen (RWTH), Aachen, Germany.
J Biomater Appl. 2011 May;25(8):771-93. doi: 10.1177/0885328209360423. Epub 2010 Mar 17.
Mesh implants are frequently used in congenital diaphragmatic hernia. This experimental study aimed to examine the influence of different materials on the diaphragmatic movement over time as well as their mechanical qualities after 4 months. Ultrapro®, Surgisis®, and Proceed® were implanted onto a diaphragmatic defect in growing rabbits. Diaphragmatic mobility was determined at three time points. At 4 months, defect shrinkage and mechanical properties were measured. The break strength decreased for Ultrapro® and Surgisis®, but did not change relevantly for Proceed®. Ultrapro® (32.46 N/cm) and Proceed® (31.75 N/cm) showed a four-fold higher resistance to tearing than Surgisis® (8.31 N/cm). The elasticity of Ultrapro® showed no significant difference compared to Surgisis® (p = 0.75). Proceed®, on the other hand, was more than twice as elastic as Ultrapro® or Surgisis® (p = 0.015). Ultrapro® had a higher spring rate (6.48 N/mm) compared to Surgisis® (3.82 N/mm) or Proceed® (5.23 N/mm). Observing the standardized movement rates of the diaphragm for each mesh group over time the only statistical differences were seen for the Proceed® group. On account of its material qualities Ultrapro® was found to be the most suitable mesh material for demanding locations in our model.
网片在先天性膈疝中经常被使用。本实验研究旨在研究不同材料对随时间推移的膈运动的影响以及 4 个月后它们的机械性能。Ultrapro®、Surgisis®和 Proceed®被植入到生长中的兔子的膈缺损中。在三个时间点测定膈的活动性。在 4 个月时,测量缺损的缩小和机械性能。Ultrapro®和 Surgisis®的断裂强度下降,但 Proceed®没有明显变化。Ultrapro®(32.46 N/cm)和 Proceed®(31.75 N/cm)撕裂阻力比 Surgisis®(8.31 N/cm)高四倍。Ultrapro®的弹性与 Surgisis®相比无显著差异(p = 0.75)。另一方面,Proceed®的弹性比 Ultrapro®或 Surgisis®高两倍以上(p = 0.015)。Ultrapro®的弹簧率(6.48 N/mm)高于 Surgisis®(3.82 N/mm)或 Proceed®(5.23 N/mm)。观察每个网片组的膈标准化运动率随时间的变化,仅在 Proceed®组中观察到统计学差异。就其材料质量而言,Ultrapro®被发现是我们模型中高要求部位最适合的网片材料。