Pediatr Res. 2010 Jun;67(6):624-9. doi: 10.1203/PDR.0b013e3181dcd580.
Bubble continuous positive airway pressure (B-CPAP) applies small-amplitude, high-frequency oscillations in airway pressure (DeltaPaw) that may improve gas exchange in infants with respiratory disease. We developed a device, high-amplitude B-CPAP (HAB-CPAP), which provides greater DeltaPaw than B-CPAP provides. We studied the effects of different operational parameters on DeltaPaw and volumes of gas delivered to a mechanical infant lung model. In vivo studies tested the hypothesis that HAB-CPAP provides noninvasive respiratory support greater than that provided by B-CPAP. Lavaged juvenile rabbits were stabilized on ventilator nasal CPAP. The animals were then supported at the same mean airway pressure, bias flow, and fraction of inspired oxygen (FiO2) required for stabilization, whereas the bubbler angle was varied in a randomized crossover design at exit angles, relative to vertical, of 0 (HAB-CPAP0; equivalent to conventional B-CPAP), 90 (HAB-CPAP90), and 135 degrees (HAB-CPAP135). Arterial blood gases and pressure-rate product (PRP) were measured after 15 min at each bubbler angle. Pao2 levels were higher (p<0.007) with HAB-CPAP135 than with conventional B-CPAP. PaCO2 levels did not differ (p=0.073) among the three bubbler configurations. PRP with HAB-CPAP135 were half of the PRP with HAB-CPAP0 or HAB-CPAP90 (p=0.001). These results indicate that HAB-CPAP135 provides greater respiratory support than conventional B-CPAP does.
气泡持续气道正压通气(B-CPAP)施加小振幅、高频率的气道压力振荡(DeltaPaw),可能改善呼吸疾病婴儿的气体交换。我们开发了一种设备,高振幅 B-CPAP(HAB-CPAP),它提供比 B-CPAP 更大的 DeltaPaw。我们研究了不同操作参数对机械婴儿肺模型输送的气体的 DeltaPaw 和体积的影响。体内研究检验了以下假设:HAB-CPAP 提供的无创呼吸支持大于 B-CPAP 提供的支持。冲洗的幼兔在呼吸机上进行鼻 CPAP 稳定。然后,以稳定所需的相同平均气道压力、偏流和吸入氧分数(FiO2)支持动物,而在出口角度以相对于垂直的 0(HAB-CPAP0;相当于常规 B-CPAP)、90(HAB-CPAP90)和 135 度(HAB-CPAP135)的随机交叉设计改变 bubbler 角度。在每个 bubbler 角度下,15 分钟后测量动脉血气和压力-速率产物(PRP)。与常规 B-CPAP 相比,HAB-CPAP135 的 Pao2 水平更高(p<0.007)。三种 bubbler 配置之间的 PaCO2 水平没有差异(p=0.073)。HAB-CPAP135 的 PRP 是 HAB-CPAP0 或 HAB-CPAP90 的 PRP 的一半(p=0.001)。这些结果表明,HAB-CPAP135 提供比常规 B-CPAP 更大的呼吸支持。