Nève J, Leclercq N
Department of Pharmaceutical Organic Chemistry, Free University of Brussels, Belgium.
Clin Chem. 1991 May;37(5):723-8.
A method was developed for the routine determination of manganese in serum from healthy human subjects by graphite furnace atomic absorption spectrometry. We controlled the experimental conditions rigorously, from sampling to analysis, to minimize contamination and to conserve diluted samples. We optimized the procedure with two serum Reference Materials, one of which had a manganese concentration very close to what is thought to be the physiological concentration in humans. The best analytical performance was obtained by directly injecting into the furnace serum diluted with an equal volume of a solution containing Triton X-100 and sodium EDTA and calibrating by the standard additions procedure or by a calibration graph constructed in a similar matrix. The Zeeman background correction produced better accuracy and precision than did the classical deuterium correction. Within-run CV for a manganese concentration of 12.7 nmol/L in serum was 7.9%, and between-run precision was 16.1%. The mean (SD) serum manganese concentration in 31 healthy adults was 10.8 (SD 3.0) nmol/L. Sex and age of subjects did not affect concentrations.
开发了一种通过石墨炉原子吸收光谱法常规测定健康人体血清中锰的方法。从采样到分析,我们严格控制实验条件,以尽量减少污染并保存稀释后的样品。我们用两种血清标准物质对该方法进行了优化,其中一种的锰浓度非常接近人们认为的人体生理浓度。通过将等体积含有 Triton X - 100 和乙二胺四乙酸钠的溶液稀释后的血清直接注入炉中,并采用标准加入法或在类似基质中构建的校准曲线进行校准,可获得最佳分析性能。与经典的氘校正相比,塞曼背景校正具有更高的准确度和精密度。血清中锰浓度为 12.7 nmol/L 时,批内变异系数为 7.9%,批间精密度为 16.1%。31 名健康成年人的血清锰平均(标准差)浓度为 10.8(标准差 3.0)nmol/L。受试者的性别和年龄不影响浓度。