评价 Cadi ThermoSENSOR 无线皮肤接触体温计在儿童测量耳温和腋温的准确性。

Evaluation of the Cadi ThermoSENSOR wireless skin-contact thermometer against ear and axillary temperatures in children.

机构信息

Upper Iowa University, Singapore.

出版信息

J Pediatr Nurs. 2010 Jun;25(3):176-86. doi: 10.1016/j.pedn.2008.12.002. Epub 2009 Apr 18.

Abstract

The Cadi ThermoSENSOR skin-contact thermometer measures body temperature continuously and transmits readings wirelessly to a central server. This study evaluated the ThermoSENSOR against ear temperatures (ETs) measured by a Braun ThermoScan ear thermometer and axillary temperatures (ATs) measured by a Terumo digital clinical thermometer. The test participants consisted of 109 children aged 6 months to 16 years from a pediatric ward. The sensor was attached to the lower abdomen at least 15 minutes before the first measurement. ET, AT, and ThermoSENSOR temperatures (TTs) were recorded up to three times at the usual measurement times. The TTs differed from ETs by -0.23 degrees C +/- 0.47 degrees C (mean +/- standard deviation, n = 271) and from ATs by +0.21 degrees C +/- 0.46 degrees C (n = 270). The ETs differed from ATs by +0.43 degrees C +/- 0.42 degrees C (n = 315). These results suggest that the TTs were comparable to the ETs and ATs.

摘要

Cadi ThermoSENSOR 皮肤接触式体温计持续测量体温,并将读数无线传输到中央服务器。本研究评估了 ThermoSENSOR 与 Braun ThermoScan 耳温计测量的耳温 (ET) 和 Terumo 数字临床体温计测量的腋温 (AT) 的对比。试验参与者为儿科病房的 109 名 6 个月至 16 岁的儿童。传感器至少在第一次测量前 15 分钟贴在下腹部。在常规测量时间内,最多记录 3 次 ET、AT 和 ThermoSENSOR 温度 (TT)。TT 与 ET 的差值为 -0.23 摄氏度 +/- 0.47 摄氏度(n = 271),与 AT 的差值为 +0.21 摄氏度 +/- 0.46 摄氏度(n = 270)。ET 与 AT 的差值为 +0.43 摄氏度 +/- 0.42 摄氏度(n = 315)。这些结果表明 TT 与 ET 和 AT 相当。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/ef06/7127543/96e77e810877/gr1.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验