国际多中心随机对照研究比较鼻电位差技术。

An international randomized multicenter comparison of nasal potential difference techniques.

机构信息

University of Alabama at Birmingham, Birmingham, AL 35294-0006, USA.

出版信息

Chest. 2010 Oct;138(4):919-28. doi: 10.1378/chest.10-0179. Epub 2010 May 14.

Abstract

BACKGROUND

The transepithelial nasal potential difference (NPD) is used to assess cystic fibrosis transmembrane conductance regulator (CFTR) activity. Unreliability, excessive artifacts, and lack of standardization of current testing systems can compromise its use as a diagnostic test and outcome measure for clinical trials.

METHODS

To determine whether a nonperfusing (agar gel) nasal catheter for NPD measurement is more reliable and less susceptible to artifacts than a continuously perfusing nasal catheter, we performed a multicenter, randomized, crossover trial comparing a standardized NPD protocol using an agar nasal catheter with the same protocol using a continuously perfusing catheter. The data capture technique was identical in both protocols. A total of 26 normal adult subjects underwent NPD testing at six different centers.

RESULTS

Artifact frequency was reduced by 75% (P < .001), and duration was less pronounced using the agar catheter. The measurement of sodium conductance was similar between the two catheter methods, but the agar catheter demonstrated significantly greater CFTR-dependent hyperpolarization, because Δ zero Cl- + isoproterenol measurements were significantly more hyperpolarized with the agar catheter (224.2 ± 12.9 mV with agar vs 18.2 ± 9.1 mV with perfusion, P < .05).

CONCLUSIONS

The agar nasal catheter approach demonstrates superior reliability compared with the perfusion nasal catheter method for measurement of NPD. This nonperfusion catheter method should be considered for adoption as a standardized protocol to monitor CFTR activity in clinical trials.

摘要

背景

跨上皮鼻电位差 (NPD) 用于评估囊性纤维化跨膜电导调节剂 (CFTR) 活性。当前测试系统的不可靠性、过多的伪影以及缺乏标准化会影响其作为诊断测试和临床试验结果衡量标准的应用。

方法

为了确定非灌注(琼脂凝胶)鼻导管测量 NPD 是否比连续灌注鼻导管更可靠且不易产生伪影,我们进行了一项多中心、随机、交叉试验,比较了使用琼脂鼻导管的标准化 NPD 方案与使用连续灌注导管的相同方案。两种方案中的数据捕获技术完全相同。共有 26 名正常成年受试者在六个不同中心接受 NPD 测试。

结果

使用琼脂导管可将伪影频率降低 75%(P <.001),且持续时间更短。两种导管方法的钠电导测量值相似,但琼脂导管显示出更显著的 CFTR 依赖性超极化,因为 Δ zero Cl- + 异丙肾上腺素测量值用琼脂导管时更明显超极化(琼脂导管为 224.2 ± 12.9 mV,灌注导管为 18.2 ± 9.1 mV,P <.05)。

结论

与灌注鼻导管方法相比,琼脂鼻导管方法在测量 NPD 方面具有更高的可靠性。这种非灌注导管方法应被视为监测临床试验中 CFTR 活性的标准化方案。

相似文献

引用本文的文献

10
Use of ferrets for electrophysiologic monitoring of ion transport.使用雪貂进行离子转运的电生理监测。
PLoS One. 2017 Oct 27;12(10):e0186984. doi: 10.1371/journal.pone.0186984. eCollection 2017.

本文引用的文献

9
Cystic fibrosis.囊性纤维化
N Engl J Med. 2005 May 12;352(19):1992-2001. doi: 10.1056/NEJMra043184.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验