Evaluation of effect of specimen-handling parameters for plasma preparation tubes on viral load measurements obtained by using the Abbott RealTime HIV-1 load assay.

机构信息

Department of Pathology, University of Medicine and Dentistry of New Jersey, New Jersey Medical School, Newark, NJ 07103, USA.

出版信息

J Clin Microbiol. 2010 Jul;48(7):2464-8. doi: 10.1128/JCM.00083-10. Epub 2010 May 19.

Abstract

HIV-1 viral load testing is essential to the management of HIV-1-infected patients, and proper specimen handling ensures accurate viral load (VL) results. This study was performed to (i) evaluate the effect of freezing plasma in situ in BD Vacutainer plasma preparation tubes (PPT) on the accuracy of HIV-1 viral load results using the Abbott RealTime HIV-1 assay and (ii) evaluate the effect of whole-blood storage in the PPT for 6 h at room temperature prior to centrifugation (PPT6H) rather than 2 h as specified in the PPT product insert. Of the 64 HIV-positive subjects evaluated, 29 had average viral load counts of >40 copies/ml in at least one of the tubes tested and 35 subjects had a result of either "undetected target" or "below the limit of quantification" (LOQ) for all or some of the tubes regardless of handling condition. For the 29 subjects with VLs that were >LOQ, the mean biases between plasma from Vacutainer K(2)EDTA tubes and plasma frozen in situ in PPT and between K(2)EDTA tube plasma and plasma from PPT6H (log(10) copies/ml) were 0.005 and -0.001, respectively, and r(2) was >0.92 for all correlations. We conclude that VLs determined from plasma frozen in situ in PPT are equivalent to VLs in K(2)EDTA tube plasma and can be used for accurate quantification of HIV-1 RNA in the Abbott RealTime HIV-1 assay. Furthermore specimens collected in PPT can be stored for 6 h at room temperature with no effect on viral load results as measured by the Abbott RealTime HIV-1 assay.

摘要

HIV-1 病毒载量检测对于 HIV-1 感染患者的管理至关重要,适当的标本处理可确保准确的病毒载量 (VL) 结果。本研究旨在:(i) 评估使用 Abbott RealTime HIV-1 检测时,将血浆直接冷冻于 BD Vacutainer 血浆制备管 (PPT) 中对 HIV-1 病毒载量结果准确性的影响,以及 (ii) 评估室温下在 PPT 中储存全血 6 小时(而非 PPT 产品说明书中规定的 2 小时),再离心对结果的影响。在 64 例 HIV 阳性的受试者中,29 例至少有一个管的平均病毒载量计数 >40 拷贝/ml,35 例受试者所有或部分管的结果均为“未检测到靶标”或“低于定量下限 (LOQ)”,而与处理条件无关。对于 29 例 VL 高于 LOQ 的受试者,从 Vacutainer K(2)EDTA 管血浆和直接在 PPT 中冷冻的血浆以及从 K(2)EDTA 管血浆和 PPT6H 血浆(log(10) 拷贝/ml)之间的平均偏差分别为 0.005 和 -0.001,所有相关性的 r(2) 值均 >0.92。我们得出结论,直接在 PPT 中冷冻的血浆的 VL 与 K(2)EDTA 管血浆的 VL 相当,可用于 Abbott RealTime HIV-1 检测中 HIV-1 RNA 的准确定量。此外,在 Abbott RealTime HIV-1 检测中,使用 PPT 采集的标本可在室温下储存 6 小时,对病毒载量结果无影响。

相似文献

本文引用的文献

6
HIV-1 load quantitation: a 17-year perspective.HIV-1载量定量:17年的视角
J Infect Dis. 2006 Sep 15;194 Suppl 1:S38-44. doi: 10.1086/505352.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验