Cross-axis synchronous flow-through coil planet centrifuge (type XLL). I. Design of the apparatus and studies on retention of stationary phase.

作者信息

Ito Y, Kitazume E, Bhatnagar M, Trimble F D

机构信息

Laboratory of Biophysical Chemistry, National Heart, Lung, and Blood Institute, National Institutes of Health, Bethesda, MD 20892.

出版信息

J Chromatogr. 1991 Jan 18;538(1):59-66. doi: 10.1016/s0021-9673(01)91621-8.

Abstract

The fourth prototype holds a pair of column holders in the lateral position at 15 cm from the center of the rotary shaft horizontally mounted on the rotary frame at 7.6 cm from the central axis of the apparatus. Using short coils of 2.6 mm I.D. PTFE (polytetrafluoroethylene) tubing with 7.6 cm and 24 cm helical diameters, retention of the stationary phase was measured in ten pairs of two-phase solvent systems under various experimental conditions. Satisfactory retention was obtained by choosing proper combinations of three factors, i.e., the direction of planetary motion, head-tail elution mode, and inward-outward elution mode. The polar butanol solvent systems showed excellent retention from 65 to 80% in the 7.6 cm helical diameter left-handed coil.