To determine the effect of suture tension on contact pressures in the lateral compartment of the stifle for the TightRope CCL((R)) (TR) and circumfabellar-tibial suture (LS) techniques.
Ex vivo biomechanical study.
Pelvic limbs (n=12) from dogs weighing 21-24 kg.
A pressure sensor was used to record peak contact pressure (PCP) and mean contact pressure (MCP) and contact area (CA) in limbs mounted in a materials testing machine. Stifles were stabilized using either a TR or LS technique and the prosthesis was tensioned increasingly from 25 to 70 N. PCP, MCP, and CA were collected at each applied tension before and after applying an axial load of 150 N.
In the unloaded condition, LS stabilization resulted in a significant increase (P=.034) in PCP at >or=40 N tensions, whereas TR stabilization resulted in a significant increase (P=.047) in PCP at 70 N tension. There was a significant increase in CA at tensions >or=60 N in the LS group. No effect on MCP was noted. No significant differences in contact pressures were observed in the loaded condition. There were no significant differences in normalized PCP and CA between stabilization groups at any of the tension increment.
The application of high tension to the extra-articular prosthesis alters lateral compartment contact pressures in the unloaded stifle in this ex vivo model.
Over tightening of an extra-articular prosthesis may increase lateral compartment pressure at the time of surgery. The clinical implications of this finding are unknown, as weight bearing may negate any detrimental effect of high prosthesis tension.
确定缝线张力对用于治疗犬后十字韧带(CCL)损伤的TightRope CCL(TR)和环腓骨-胫骨缝线(LS)技术在膝关节外侧间室接触压力的影响。
体外生物力学研究。
体重21 - 24千克犬的骨盆肢(n = 12)。
使用压力传感器记录安装在材料试验机中的肢体的峰值接触压力(PCP)、平均接触压力(MCP)和接触面积(CA)。使用TR或LS技术稳定膝关节,并将假体张力从25 N逐渐增加到70 N。在施加150 N轴向载荷之前和之后,在每个施加的张力下收集PCP、MCP和CA。
在无载荷条件下,LS稳定化在张力≥40 N时导致PCP显著增加(P = 0.034),而TR稳定化在70 N张力时导致PCP显著增加(P = 0.047)。在LS组中,张力≥60 N时CA显著增加。未观察到对MCP有影响。在加载条件下,未观察到接触压力有显著差异。在任何张力增量下,稳定化组之间的标准化PCP和CA均无显著差异。
在这个体外模型中,对关节外假体施加高张力会改变无载荷膝关节外侧间室的接触压力。
关节外假体过度收紧可能在手术时增加外侧间室压力。由于负重可能抵消高假体张力的任何有害影响,这一发现的临床意义尚不清楚。