Boehm Andreas G, Weber Anja, Pillunat Lutz E, Koch Rainer, Spoerl Eberhard
Department of Ophthalmology, University of Dresden, Dresden, Germany.
Invest Ophthalmol Vis Sci. 2008 Jun;49(6):2472-7. doi: 10.1167/iovs.07-1366. Epub 2008 Mar 3.
The purpose of the study was to examine the effect of central corneal thickness (CCT), corneal curvature, astigmatism, axial length (AL), and age on measurements with the Pascal Dynamic Contour Tonometer (DCT).
In a prospective clinical trial 75 eyes of 75 patients undergoing phacoemulsification were examined. Before phacoemulsification, the anterior chamber was cannulated at the temporal corneal limbus. In a closed system the IOP was directly set to 15, 20, or 35 mm Hg with a manometric water column. IOP measurements taken by DCT were compared to intracameral measurements with a precision reference pressure sensor.
Measurements from 60 patients were suitable for statistical analysis. At IOP of 15 mm Hg, the mean difference between IOP measured by DCT and intracameral IOP was -0.02 +/- 1.32 mm Hg; at 20 mm Hg it was -0.2 +/- 1.44 mm Hg and at 35 mm Hg, -0.84 +/- 1.90 mm Hg. The concordance coefficient according to Lin was 0.9763, showing good agreement between DCT- and intracamerally measured IOP. There was a statistically significant correlation between the difference in IOP measured by DCT minus intracameral IOP and CCT (P = 0.0291, R2 = 0.00012). All other parameters had no statistically significant effect on the difference between DCT and intracameral IOP (corneal curvature, P = 0.6094, R2 = 0.00367; age, P = 0.9198, R2 = 0.000003; astigmatism, P = 0.1564, R2 = 0.08497; and axial length, P = 0.9484, R2 = 0.00008).
Measurements with the DCT showed good concordance with intracameral IOP. CCT exerted a statistically significant but clinically irrelevant effect on measurements with the DCT.
本研究旨在探讨中央角膜厚度(CCT)、角膜曲率、散光、眼轴长度(AL)和年龄对帕斯卡动态轮廓眼压计(DCT)测量结果的影响。
在一项前瞻性临床试验中,对75例接受超声乳化手术患者的75只眼进行了检查。在超声乳化手术前,于颞侧角膜缘穿刺前房。在封闭系统中,通过压力水柱将眼压直接设定为15、20或35 mmHg。将DCT测量的眼压与使用精密参考压力传感器进行的前房内测量结果进行比较。
60例患者的测量结果适合进行统计分析。眼压为15 mmHg时,DCT测量的眼压与前房内眼压的平均差值为-0.02±1.32 mmHg;眼压为20 mmHg时,差值为-0.2±1.44 mmHg;眼压为35 mmHg时,差值为-0.84±1.90 mmHg。根据林氏法得出的一致性系数为0.9763,表明DCT测量的眼压与前房内测量的眼压之间具有良好的一致性。DCT测量的眼压减去前房内眼压的差值与CCT之间存在统计学显著相关性(P = 0.0291,R2 = 0.00012)。所有其他参数对DCT与前房内眼压的差值均无统计学显著影响(角膜曲率,P = 0.6094,R2 = 0.00367;年龄,P = 0.9198,R2 = 0.000003;散光,P = 0.1564,R2 = 0.08497;眼轴长度,P = 0.9484,R2 = 0.00008)。
DCT测量结果与前房内眼压具有良好的一致性。CCT对DCT测量结果有统计学显著但临床意义不大的影响。