热电偶与电子温度计连接时在 5 种水浴温度下的响应。

Response of thermocouples interfaced to electrothermometers when immersed in 5 water bath temperatures.

机构信息

Department of Health and Human Performance, Oklahoma State University, Stillwater, USA.

出版信息

J Athl Train. 2010 Jul-Aug;45(4):338-43. doi: 10.4085/1062-6050-45.4.338.

Abstract

CONTEXT

Thermocouples and electrothermometers are used in therapeutic modality research. Until recently, researchers assumed that these instruments were valid and reliable.

OBJECTIVE

To examine 3 different thermocouple types in 5 degrees C, 15 degrees C, 18.4 degrees C, 25 degrees C, and 35 degrees C water baths.

DESIGN

Randomized controlled trial.

SETTING

Therapeutic modality laboratory.

INTERVENTION(S): Eighteen thermocouple leads were inserted through the wall of a foamed polystyrene cooler. The cooler was filled with water. Six thermocouples (2 of each model) were plugged into the 6 channels of the Datalogger and 6 randomly selected channels in the 2 Iso-Thermexes. A mercury thermometer was immersed into the water and was read every 10 seconds for 4 minutes during each of 6 trials. The entire process was repeated for each of 5 water bath temperatures (5 degrees C, 15 degrees C, 18.4 degrees C, 25 degrees C, 35 degrees C).

MAIN OUTCOME MEASURE(S): Temperature and absolute temperature differences among 3 thermocouple types (IT-21, IT-18, PT-6) and 3 electrothermometers (Datalogger, Iso-Thermex calibrated from -50 degrees C to 50 degrees C, Iso-Thermex calibrated from -20 degrees C to 80 degrees C).

RESULTS

Validity and reliability were dependent on thermocouple type, electrothermometer, and water bath temperature (P < .001; modified Levene P < .05). Statistically, the IT-18 and PT-6 thermocouples were not reliable in each electrothermometer; however, these differences were not practically different from each other. The PT-6 thermocouples were more valid than the IT-18s, and both thermocouple types were more valid than the IT-21s, regardless of water bath temperature (P < .001).

CONCLUSIONS

The validity and reliability of thermocouples interfaced to an electrothermometer under experimental conditions should be tested before data collection. We also recommend that investigators report the validity, the reliability, and the calculated uncertainty (validity + reliability) of their temperature measurements for therapeutic modalities research. With this information, investigators and clinicians will be better able to interpret and compare results and conclusions.

摘要

背景

热电偶和电子体温计在治疗方式研究中被广泛使用。直到最近,研究人员都假定这些仪器是有效且可靠的。

目的

在 5°C、15°C、18.4°C、25°C 和 35°C 的水浴中,检查 3 种不同类型的热电偶。

设计

随机对照试验。

设置

治疗方式实验室。

干预

将 18 个热电偶探头插入泡沫聚苯乙烯冷却器的壁中。冷却器充满水。将 6 个热电偶(每种型号 2 个)插入数据记录器的 6 个通道和 2 个 Iso-Thermex 中的 6 个随机选择的通道。将一个水银温度计浸入水中,在 6 次试验的每一次中,每隔 10 秒读取 4 分钟的温度。在每个水浴温度(5°C、15°C、18.4°C、25°C、35°C)下,整个过程都重复进行。

主要观察指标

3 种热电偶类型(IT-21、IT-18、PT-6)和 3 种电子体温计(Datalogger、从-50°C 到 50°C 校准的 Iso-Thermex、从-20°C 到 80°C 校准的 Iso-Thermex)之间的温度和绝对温度差异。

结果

有效性和可靠性取决于热电偶类型、电子体温计和水浴温度(P<.001;经修正的 Levene P<.05)。统计上,在每个电子体温计中,IT-18 和 PT-6 热电偶均不可靠;然而,彼此之间的差异在实际应用中并无明显不同。PT-6 热电偶比 IT-18 热电偶更准确,且在任何水浴温度下,两种热电偶类型都比 IT-21 热电偶更准确(P<.001)。

结论

在收集数据之前,应在实验条件下测试与电子体温计接口的热电偶的有效性和可靠性。我们还建议研究人员报告治疗方式研究中他们的温度测量的有效性、可靠性和计算出的不确定性(有效性+可靠性)。有了这些信息,研究人员和临床医生将能够更好地解释和比较结果和结论。